KANE3500 Manual
Page 8
Always use the top right hand port (Port 1) for taking single channel measurements
(
pressure or vacuum). Only use the left hand port for differential measurements.
If the pressure being measured exceeds the meter’s design range the display will show
OR
for “over-range”
When taking draft readings at very low pressure or draft levels, for maximum accuracy, re-
zero the meter in the orientation that it is being used. This eliminated gravity effects on the
very sensitive pressure transducer. It is also recommend that the meter is switched on for at
least five minutes and then re-zeroed before taking such sensitive measurements.
EXAMPLE PRINTOUTS
The standard printout is:
The MAX and MIN readings are those observed since last switch on
NOTE:
Printouts of stored readings will also include the TEST NO. below the header.