O P E R AT I O N C O N T I N U E D
FIG. 4-5
Output data OOD models serial numbers OOD-19-141 or higher
Column 1
– Message ID (for information Only)
Column 2
– Time code (for information only)
Column 3
– Background Signal: Shows the output value from the PD sensor when the light
source is off. Range 0 to 2,500
Column 4
– PD sensor signal: Shows the output value from the PD sensor when the
light source is on. Range background sensor signal value to 65,535
Column 5
– LED Off signal: Shows the value of the LED light source when it is off.
Range: 0 to 150
Column 6
– LED On signal: Shows the value of the LED light source when it is
on. Range: 500 to 1,023
Column 7
– PD sensor voltage: Range 47.0 to 55.0
Column 8
– Differential value between the PD sensor signal
and the background signal
Column 9
– Light source compensated PD sensor signal
Column 10
– Offset level compensated PD sensor
signal
Column 11
– Oil Output: Range 0 to 5,000
ppm (Max)
Column 12
– 4-20 mA Output Loop
Column 13
– Electronics
temperature. Range: 0.0 to
70.0 °C
Column 14
- CRC
Cyclic Redundancy
Check (for information
only)