Optidew User’s Manual
48
October 2019
APPENDIX B
Appendix B
Modbus Register Map
All the data values relating to the Optidew are stored in 16-bit wide holding registers.
Registers can contain either measured or calculated values (dewpoint, temperature,
relative humidity etc.), or configuration data (analog output or alarm settings).
Modbus RTU Implementation
This is a partial implementation of the Modbus RTU Standard with the following codes
implemented:
Function Code
Description
3
Read Holding Register
6
Write Holding Register
16
Write Multiple Holding Registers
Register Types
Data Type
Description
float
IEE754 32 bit single precision floating point, spans 2 16-bit
holding registers. First register contains the most significant bits.
uint16
16 bit unsigned integer, can contain options list e.g. 0 = Dew
Point, 1 = Temperature.
int16
16 bit signed integer.
boolean
Can be treated like a uint16, where 0 = false/disabled, and 1 is
true/enabled.
Communications
In order to communicate with the instrument over a USB connection, first install the
Kahn application software which contains a USB->UART bridge driver. The Optidew will
then appear in Device Manager as a virtual serial port.
Serial Port Settings (USB/RS485)
9600 Baud Rate, 8 Data Bits, No Parity, 1 Stop Bit, No Flow Control
Modbus TCP
If using the Ethernet connection, the instrument uses the Modbus TCP protocol instead
of Modbus RTU. Refer to resources online for the key differences.
Summary of Contents for Optidew Bench Top
Page 2: ...Optidew For Kahn Instruments contact information please go to www kahn com...
Page 13: ...Optidew User s Manual 4 October 2019 INSTALLATION This page is intentially left blank...
Page 53: ...Optidew User s Manual 44 October 2019 APPENDIX A Appendix A Technical Specifications...
Page 56: ...Optidew User s Manual Kahn Instruments 47 APPENDIX B Appendix B Modbus Register Map...
Page 65: ...Optidew User s Manual 56 October 2019 APPENDIX C Appendix C Dimensional Drawings...
Page 70: ...Optidew User s Manual Kahn Instruments 61 APPENDIX C Dual Stage Dewpoint Sensor...
Page 71: ...Optidew User s Manual 62 October 2019 APPENDIX C Sample Block...
Page 74: ...Optidew User s Manual Kahn Instruments 65 APPENDIX C...
Page 75: ...Optidew User s Manual 66 October 2019 APPENDIX C...
Page 76: ...Optidew User s Manual Kahn Instruments 67 APPENDIX C...
Page 77: ...Optidew User s Manual 68 October 2019 APPENDIX C...
Page 78: ...Optidew User s Manual Kahn Instruments 69 APPENDIX C...
Page 79: ...Optidew User s Manual 70 October 2019 APPENDIX C...
Page 80: ...Optidew User s Manual Kahn Instruments 71 APPENDIX C...
Page 81: ...Optidew User s Manual 72 October 2019 APPENDIX C...
Page 82: ...Optidew User s Manual Kahn Instruments 73 APPENDIX C...
Page 83: ...Optidew User s Manual 74 October 2019 APPENDIX C...
Page 84: ...Optidew User s Manual Kahn Instruments 75 APPENDIX C...
Page 85: ...Optidew User s Manual 76 October 2019 APPENDIX C...
Page 86: ...Optidew User s Manual Kahn Instruments 77 APPENDIX C...
Page 87: ...Optidew User s Manual 78 October 2019 APPENDIX C...
Page 88: ...Optidew User s Manual Kahn Instruments 79 APPENDIX C...