Optidew User’s Manual
30
October 2019
OPERATION
4
OPERATION
4.1
Operating Cycle
Figure 24
Typical operating cycle
At initial switch-on, the instrument enters a DCC cycle for 2 minutes. During this time the
mirror is heated above the prevailing dewpoint to ensure that all condensate is driven
off the surface of the mirror. The degree of heating is determined by the configuration
of the ‘Type’ and ‘Setpoint’ parameters in the DCC menu (see section 3.2 for further
information).
The mirror is maintained at this temperature for the DCC period (default 4 minutes) or
2 minutes on switch-on. During the DCC process, Data Hold fixes the analog outputs
at the same value(s) as before DCC commenced. Data Hold typically lasts 4 minutes
from the end of a DCC cycle, or until the instrument has reached the dewpoint. This
procedure is in place to prevent any system which is connected to the outputs from
receiving a 'false' reading.
After the DCC period has finished, the measurement period commences, during which
the control system decreases the mirror temperature until it reaches the dewpoint. The
sensor will take a short amount of time to form a film of condensate and control on the
dewpoint. The length of this stabilization time depends upon the dewpoint temperature.
When the measurement is stable or tracking very slow changes in dewpoint, the Sensor
indicator in the Operational Status display will indicate ‘Control’. Note that at dry
dewpoints (below around -4°F/-20°C) the sensor may display ‘Control’ when the mirror
temperature is still slowly oscillating, always use the trend graph on the display as a
secondary indication.
The end of a DCC cycle re-sets the interval counter, meaning that another DCC will start
(by default) after 4 hours have elapsed. Once the measurement is stable, Data Hold will
release, and the analog outputs will resume their normal operation. At this point the
Status area of the Operational Status display will change to ‘Measure’.
Summary of Contents for Optidew Bench Top
Page 2: ...Optidew For Kahn Instruments contact information please go to www kahn com...
Page 13: ...Optidew User s Manual 4 October 2019 INSTALLATION This page is intentially left blank...
Page 53: ...Optidew User s Manual 44 October 2019 APPENDIX A Appendix A Technical Specifications...
Page 56: ...Optidew User s Manual Kahn Instruments 47 APPENDIX B Appendix B Modbus Register Map...
Page 65: ...Optidew User s Manual 56 October 2019 APPENDIX C Appendix C Dimensional Drawings...
Page 70: ...Optidew User s Manual Kahn Instruments 61 APPENDIX C Dual Stage Dewpoint Sensor...
Page 71: ...Optidew User s Manual 62 October 2019 APPENDIX C Sample Block...
Page 74: ...Optidew User s Manual Kahn Instruments 65 APPENDIX C...
Page 75: ...Optidew User s Manual 66 October 2019 APPENDIX C...
Page 76: ...Optidew User s Manual Kahn Instruments 67 APPENDIX C...
Page 77: ...Optidew User s Manual 68 October 2019 APPENDIX C...
Page 78: ...Optidew User s Manual Kahn Instruments 69 APPENDIX C...
Page 79: ...Optidew User s Manual 70 October 2019 APPENDIX C...
Page 80: ...Optidew User s Manual Kahn Instruments 71 APPENDIX C...
Page 81: ...Optidew User s Manual 72 October 2019 APPENDIX C...
Page 82: ...Optidew User s Manual Kahn Instruments 73 APPENDIX C...
Page 83: ...Optidew User s Manual 74 October 2019 APPENDIX C...
Page 84: ...Optidew User s Manual Kahn Instruments 75 APPENDIX C...
Page 85: ...Optidew User s Manual 76 October 2019 APPENDIX C...
Page 86: ...Optidew User s Manual Kahn Instruments 77 APPENDIX C...
Page 87: ...Optidew User s Manual 78 October 2019 APPENDIX C...
Page 88: ...Optidew User s Manual Kahn Instruments 79 APPENDIX C...