Optidew User’s Manual
Kahn Instruments
31
OPERATION
4.2
Operating Guide
4.2.1 Description
Once the Optidew has been powered on and has carried out its’ initial DCC, it will attempt
to find the dewpoint. In order to measure the dewpoint a Chilled Mirror hygrometer
must control a thin film of condensed water or ice on the mirror.
To initially form the condensate layer the mirror must be cooled past the actual dew
or frost point. The control system will then gradually heat the mirror to reduce the
thickness of this condensate layer. It typically takes several heating/cooling cycles
until the instrument has achieved the optimal film thickness where evaporation and
condensation are occurring in equilibrium. This is the true dew/frost point of the sample.
After finding the true dewpoint, the control system will continue to maintain the film
thickness at a constant level. Any decrease in actual sample dewpoint will cause
evaporation from the condensate film to increase – reducing its thickness and causing
the control system to cool the mirror to compensate. Likewise if the dewpoint increases
then condensation on the mirror will increase, and the control system will heat to
compensate.
In extreme cases where the dewpoint decreases very abruptly, then the condensate will
be completely evaporated from the mirror. In these scenarios the system will ‘search’
for the dewpoint again by cooling, resulting in cooing past the dewpoint as described
above. A similar situation occurs when the dewpoint increases abruptly, however the
condensate film can be lost here by the control system heating to compensate and
exceeding the new dewpoint.
4.2.2 Operating Practice
There are two basic methods of measuring with the Optidew:
In-situ measurements are made by placing the sensor(s) inside the environment to be
measured.
Extractive measurements are made by installing the sensor into a block within a sample
handling system, and flowing the sample outside of the environment to be measured
through this system
Extractive measurements are recommended when the conditions in the environment
to be measured are not conducive to making reliable measurements with the product.
Examples of such conditional limitations are:
•
Excessive flow rate
•
Presence of particulates matter
•
Presence of entrained liquids
•
Excessive sample temperature
•
Dew point is beyond depression capability at sample temperature
The basic considerations for each measurement type are as follows:
In-Situ
1.
Dewpoint Sensor position
– will the sensor see an area of the
environment that is representative of what you want to measure? For
Summary of Contents for Optidew Bench Top
Page 2: ...Optidew For Kahn Instruments contact information please go to www kahn com...
Page 13: ...Optidew User s Manual 4 October 2019 INSTALLATION This page is intentially left blank...
Page 53: ...Optidew User s Manual 44 October 2019 APPENDIX A Appendix A Technical Specifications...
Page 56: ...Optidew User s Manual Kahn Instruments 47 APPENDIX B Appendix B Modbus Register Map...
Page 65: ...Optidew User s Manual 56 October 2019 APPENDIX C Appendix C Dimensional Drawings...
Page 70: ...Optidew User s Manual Kahn Instruments 61 APPENDIX C Dual Stage Dewpoint Sensor...
Page 71: ...Optidew User s Manual 62 October 2019 APPENDIX C Sample Block...
Page 74: ...Optidew User s Manual Kahn Instruments 65 APPENDIX C...
Page 75: ...Optidew User s Manual 66 October 2019 APPENDIX C...
Page 76: ...Optidew User s Manual Kahn Instruments 67 APPENDIX C...
Page 77: ...Optidew User s Manual 68 October 2019 APPENDIX C...
Page 78: ...Optidew User s Manual Kahn Instruments 69 APPENDIX C...
Page 79: ...Optidew User s Manual 70 October 2019 APPENDIX C...
Page 80: ...Optidew User s Manual Kahn Instruments 71 APPENDIX C...
Page 81: ...Optidew User s Manual 72 October 2019 APPENDIX C...
Page 82: ...Optidew User s Manual Kahn Instruments 73 APPENDIX C...
Page 83: ...Optidew User s Manual 74 October 2019 APPENDIX C...
Page 84: ...Optidew User s Manual Kahn Instruments 75 APPENDIX C...
Page 85: ...Optidew User s Manual 76 October 2019 APPENDIX C...
Page 86: ...Optidew User s Manual Kahn Instruments 77 APPENDIX C...
Page 87: ...Optidew User s Manual 78 October 2019 APPENDIX C...
Page 88: ...Optidew User s Manual Kahn Instruments 79 APPENDIX C...