background image

Easidew Plus Portable Hygrometer User’s Manual

4

  

 

97478 Issue 1, October 2014

OPERATION

OPERATION

!

Maximum operating pressure is 4350 psig (300 barg).

It is important that the gas fittings are correctly tightened, and 

any tubing is secure in the push fittings before use.

Failure to do so will affect the instrument's pressure rating.

3.1 

Instrument Gas Connections

Sample gas connections are made via the Gas in and Gas out ports located on 

the front of the instrument.  Each of these ports is equipped with  an adaptor (3) which 

screws into each gas port. 

Two types of adaptors are supplied; a small bore orifice to allow sample gas either in 

or out of the instrument and two larger bore orifices. Both types of adaptor have a ⅛” 

NPT internal thread to accommodate the customer’s choice of Gas input and Gas output 

connection fittings. 

  

A range of connection fitting options is available. 

  

Depending upon whether measurements are to be taken at atmospheric pressure

  

or line pressure, the adaptors need to be installed as shown in Figure 1A. 

 

Readings at atmospheric 

pressure

Readings at line pressure

System pressures from 15-150 

psig (1-10 barg)

Inlet:   Small Orifice (#2) Inlet: Large orifice (#1)
Outlet:  Large Orifice (#1) Outlet: Small Orifice (#2)

System pressures from 150-350 

psig (10 - 300 barg)

Large orifice (#1)

Small orifice (#2)

Figure 1A      

Adaptor Fittings

3.2 

General Operation

To operate the Easidew Plus Portable Hygrometer, follow these instructions:

1.  Insure that both the gas fittings and the Swagelok

®

 tube fittings are fully 

tightened before use.

2.  For measurement at low pressure or atmospheric pressure, no valving is 

necessary.

3.  For measurement at high pressure (150 psig to 4350 psig / 10-300 bar), 

connect the supplied metering valve to the outlet of the instrument. 

4.  Gas under test should be restricted to flow rates 2 and 10 scfh (between 

1 and 5 Nl/min) and pressures from atmospheric to 4350 psig (300 barg).

5.  Connect the sample gas supply line to the instrument 

Gas In

 port.

6.  Connect the sample gas vent line to the instrument 

Gas Out

 port.

Summary of Contents for Easidew Plus

Page 1: ...uments Inc 2015 This document is the property of Kahn Instruments and may not either in part or whole be copied or otherwise reproduced communicated in any way to third parties nor stored in any data...

Page 2: ...Easidew Plus Portable Hygrometer For Kahn Instruments contact information please go to www kahn com...

Page 3: ...14 5 4 Checking the Easidew Plus Portable Hygrometer Electronics Calibration 14 5 5 List of Spares 14 5 6 Troubleshooting 15 Contents Safety v Electrical Safety v Pressure Safety v Toxic Materials v...

Page 4: ...er supplied with the instrument are 90 to 264 V AC 47 63 Hz NOTE No other battery charger unit other than that supplied with the instrument should be used NOTE Do not allow the battery to fully discha...

Page 5: ...gram lb pound mA milliampere m meter mm millimeter MPa megapascal Nl min normal liters per minute psig pound s per square inch gauge scfh standard cubic feet per hour m micrometer ohm V Volts Warnings...

Page 6: ...meter is powered by a rechargeable NiMH battery pack giving between 12 and 16 hours of measurement time from a full charge The instrument is delivered complete with a universal battery charger stored...

Page 7: ...ccur within the case If this occurs Before opening the lid open the black pressure relief valve by one turn counter clockwise for a few seconds This will equalize the pressures When the case has been...

Page 8: ...oulder of the Swagelok tube fitting body Finger tighten the nut 5 Mark the nut at the 6 o clock position While holding the fitting body steady tighten the nut 1 turns to the 9 o clock position NOTE Co...

Page 9: ...nts are to be taken at atmospheric pressure or line pressure the adaptors need to be installed as shown in Figure 1A Readings at atmospheric pressure Readings at line pressure System pressures from 15...

Page 10: ...stem to atmospheric pressure otherwise severe injury could result NOTE Filters are essential for potentially dirty contaminated gases the installed filter should be checked before and after use and re...

Page 11: ...ument it can be accessed by removing two screws 5 The battery pack can be recharged using the supplied charger via the battery charging socket and should be recharged whenever the battery level meter...

Page 12: ...PERATION 3 5 4 20 mA Output Socket Wiring The Easidew Plus Portable Hygrometer provides a linear 4 20 mA output scaled from 76 to 140 F 60 to 60 C The socket accepts a 3 pole jack plug supplied and sh...

Page 13: ...able to water vapor as the water molecule is extremely small compared to the structure of solids even when compared to the crystalline structure of metals The graph to the right shows the dew point in...

Page 14: ...oint should always be as close to the critical measurement point as possible in order to obtain a truly representative measurement The length of the sample line to the sensor or instrument should be a...

Page 15: ...from liquid droplets and can even stop flow to the analyzer completely when a large slug of liquid is encountered Condensation and Leaks Dewpoint T Dewpoint T Maintaining the temperature of the sampl...

Page 16: ...ow the response of the sensor to changes in moisture content An excessively high flow rate can Introduce back pressure causing slower response times and unpredictable effects on equipment such as humi...

Page 17: ...pack 1 Insure that the instrument is switched off and that the battery charger current output and any sampling components are also disconnected 2 Remove and retain the 10 Phillips head screws from th...

Page 18: ...s below Completely unscrew the cable gland on the GDSN connector to release the cable tension Remove the locking screw from the back of the connector retaining the metal O ring Remove the connector bl...

Page 19: ...Optional Figure 4 2 Wire Connection View Showing Rear of Connector Terminal Block To verify if the display electronics are within calibration a 4 20mA current source can be connected in place of the...

Page 20: ...ith 4 20mA source Exchange sensor Display shows ErrH Sensor reading over range Sensor current signal 20mA Sensor element short circuit Sensor may have been exposed to saturation conditions or liquid c...

Page 21: ...Easidew Plus Portable Hygrometer User s Manual 16 97478 Issue 1 October 2014 APPENDIX A Appendix A Technical Specifications...

Page 22: ...ed Operating Conditions Operating Temperature 4 to 122 F 20 to 50 C Storage Temperature 40 to 167 F 40 to 75 C Operating Pressure Up to 4350 psig 300 barg Mechanical Specifications Display Flush mount...

Page 23: ...Easidew Plus Portable Hygrometer User s Manual 18 97478 Issue 1 October 2014 APPENDIX B Appendix B EU Declaration of Conformity...

Page 24: ...Easidew Plus Portable Hygrometer User s Manual Kahn Instruments 19 APPENDIX B Appendix B EU Declaration of Conformity...

Reviews: