Easidew Plus Portable Hygrometer User’s Manual
Kahn Instruments
3
INSTALLATION
1. Slip the nut over the
machined ferrule end of
the port connector
2. Insert the port connector
into the end connection and
finger-tighten the nut
3. While holding the fitting
body steady, tighten the
nut one-quarter turn
Do not use the Swagelok gap inspection
gauge with machined ferrule ends.
!
NOTE: Connect the machined ferrule end before
connecting the tube adapter end
4. Insert the tube adapter until it rests firmly on
the shoulder of the Swagelok
®
tube fitting body.
Finger-tighten the nut.
5. Mark the nut at the 6 o’clock position. While holding
the fitting body steady, tighten the nut 1¼ turns to the
9 o’clock position.
NOTE: Connect the machined ferrule end before
connecting the tube adapter end
4. Insert the tube adapter until it rests firmly on
the shoulder of the Swagelok
®
tube fitting body.
Finger-tighten the nut.
5. Mark the nut at the 6 o’clock position. While holding
the fitting body steady, tighten the nut 1¼ turns to the
9 o’clock position.
Figure 1
Swagelok
®
Fitting Instructions