background image

Easidew Plus Portable Hygrometer User’s Manual

Kahn Instruments

  

9

           GOOD MEASUREMENT PRACTICE

Adsorption and Desorption

Adsorption is the adhesion of atoms, ions, or molecules from a gas, liquid, or dissolved 

solid to the surface of a material, creating a film. The rate of adsorption is increased at 

higher pressures and lower temperatures.

Desorption is the release of a substance from or through the surface of a material. In 

constant environmental conditions, an adsorbed substance will remain on a surface 

almost  indefinitely.  However,  as  the  temperature  rises,  so  does  the  likelihood  of 

desorption occurring.

In practical terms, as the temperature of the environment fluctuates, water molecules 

are adsorbed and desorbed from the internal surfaces of the sample tubing, causing 

small fluctuations in the measured dew point.

Sample Tubing Length

The sample point should always be as close to the critical measurement point as possible, 

in order to obtain a truly representative measurement. The length of the sample line 

to the sensor or instrument should be as short as possible. Interconnection points and 

valves trap moisture, so using the simplest sampling arrangement possible will reduce 

the time it takes for the sample system to dry out when purged with dry gas. 

Over a long tubing run, water will inevitably migrate into any line, and the effects of 

adsorption and desorption will become more apparent. It is clear from the graph shown 

previously that the best materials to resist transpiration are stainless steel and PTFE.

Trapped Moisture

Dead volumes (areas which are not in a direct flow path) in sample lines, hold onto 

water molecules which are slowly released into the passing gas; this results in increased 

purge and response times, and wetter than expected readings. Hygroscopic materials 

in filters, valves (e.g. rubber from pressure regulators) or any other parts of the system 

can also trap moisture.

Summary of Contents for Easidew Plus

Page 1: ...uments Inc 2015 This document is the property of Kahn Instruments and may not either in part or whole be copied or otherwise reproduced communicated in any way to third parties nor stored in any data...

Page 2: ...Easidew Plus Portable Hygrometer For Kahn Instruments contact information please go to www kahn com...

Page 3: ...14 5 4 Checking the Easidew Plus Portable Hygrometer Electronics Calibration 14 5 5 List of Spares 14 5 6 Troubleshooting 15 Contents Safety v Electrical Safety v Pressure Safety v Toxic Materials v...

Page 4: ...er supplied with the instrument are 90 to 264 V AC 47 63 Hz NOTE No other battery charger unit other than that supplied with the instrument should be used NOTE Do not allow the battery to fully discha...

Page 5: ...gram lb pound mA milliampere m meter mm millimeter MPa megapascal Nl min normal liters per minute psig pound s per square inch gauge scfh standard cubic feet per hour m micrometer ohm V Volts Warnings...

Page 6: ...meter is powered by a rechargeable NiMH battery pack giving between 12 and 16 hours of measurement time from a full charge The instrument is delivered complete with a universal battery charger stored...

Page 7: ...ccur within the case If this occurs Before opening the lid open the black pressure relief valve by one turn counter clockwise for a few seconds This will equalize the pressures When the case has been...

Page 8: ...oulder of the Swagelok tube fitting body Finger tighten the nut 5 Mark the nut at the 6 o clock position While holding the fitting body steady tighten the nut 1 turns to the 9 o clock position NOTE Co...

Page 9: ...nts are to be taken at atmospheric pressure or line pressure the adaptors need to be installed as shown in Figure 1A Readings at atmospheric pressure Readings at line pressure System pressures from 15...

Page 10: ...stem to atmospheric pressure otherwise severe injury could result NOTE Filters are essential for potentially dirty contaminated gases the installed filter should be checked before and after use and re...

Page 11: ...ument it can be accessed by removing two screws 5 The battery pack can be recharged using the supplied charger via the battery charging socket and should be recharged whenever the battery level meter...

Page 12: ...PERATION 3 5 4 20 mA Output Socket Wiring The Easidew Plus Portable Hygrometer provides a linear 4 20 mA output scaled from 76 to 140 F 60 to 60 C The socket accepts a 3 pole jack plug supplied and sh...

Page 13: ...able to water vapor as the water molecule is extremely small compared to the structure of solids even when compared to the crystalline structure of metals The graph to the right shows the dew point in...

Page 14: ...oint should always be as close to the critical measurement point as possible in order to obtain a truly representative measurement The length of the sample line to the sensor or instrument should be a...

Page 15: ...from liquid droplets and can even stop flow to the analyzer completely when a large slug of liquid is encountered Condensation and Leaks Dewpoint T Dewpoint T Maintaining the temperature of the sampl...

Page 16: ...ow the response of the sensor to changes in moisture content An excessively high flow rate can Introduce back pressure causing slower response times and unpredictable effects on equipment such as humi...

Page 17: ...pack 1 Insure that the instrument is switched off and that the battery charger current output and any sampling components are also disconnected 2 Remove and retain the 10 Phillips head screws from th...

Page 18: ...s below Completely unscrew the cable gland on the GDSN connector to release the cable tension Remove the locking screw from the back of the connector retaining the metal O ring Remove the connector bl...

Page 19: ...Optional Figure 4 2 Wire Connection View Showing Rear of Connector Terminal Block To verify if the display electronics are within calibration a 4 20mA current source can be connected in place of the...

Page 20: ...ith 4 20mA source Exchange sensor Display shows ErrH Sensor reading over range Sensor current signal 20mA Sensor element short circuit Sensor may have been exposed to saturation conditions or liquid c...

Page 21: ...Easidew Plus Portable Hygrometer User s Manual 16 97478 Issue 1 October 2014 APPENDIX A Appendix A Technical Specifications...

Page 22: ...ed Operating Conditions Operating Temperature 4 to 122 F 20 to 50 C Storage Temperature 40 to 167 F 40 to 75 C Operating Pressure Up to 4350 psig 300 barg Mechanical Specifications Display Flush mount...

Page 23: ...Easidew Plus Portable Hygrometer User s Manual 18 97478 Issue 1 October 2014 APPENDIX B Appendix B EU Declaration of Conformity...

Page 24: ...Easidew Plus Portable Hygrometer User s Manual Kahn Instruments 19 APPENDIX B Appendix B EU Declaration of Conformity...

Reviews: