LOW CURRENT SETTINGS
LOW POWER SETTINGS
Low Current Set Value (Curr Undr Val):
It may bet set between (max. 50.000A) 0,1 and 50,000 A
depending on the maximum value of shunt resistance.
For ex: It may be set as max. 100A for 100A/60mV shunt.
Low Current Alarm Permit
(Curr Undr EnbL) :
YES ; alarm mod
e enabled
NO; alarm mod
e disabled
Low Current Alarm Mode Setting (Curr Undr tyPE) :
Alarm mode:
when alarm is set alarm relay is pulled. When
alarm is disabled alarm relay releases.
Latch mode:
when alarm mode is set alarm relay is pulled.
Alarm relay remains pulled even the alarm is disabled.
If latch
mode iset in models with alarm function, alarm should be
certainly disabled by pressing the set key.
Momentary mode:
In this mode, as an extra
the time seen in the adjacent box is displayed .
It may bet set between 300 msn and 5000msn.
When alarm is set, alarm relay remains pulled
until the end of this time then it is released.
This function may be used when continuous
alarm is not intended.
Low Current Delay Time (Curr Undr dELy) :
It may be set between 1 – 300 sec. If the measured current
falls below set current value, relay contact is released at the
end of the delay time.
Low Current Mistake Repair Time (Curr Undr dLy0) :
It may be set between 1 – 300 sec.
Low Current Hysteresis Value (Curr Undr
HySt) :
It may be set between % (1 – 30).
Exit from Low Current
Settings
Over Power Set Value
( Povr Over Val):
It may be set as the power value at which the relay should be
released when over power is supplied
.
Over Power Alarm Permit
(Povr Over
EnbL) :
YES ; alarm mod
e enabled
NO ; alarm mod
e disabled
Over Power Alarm Mode Setting (Povr Over tYPE) :
Alarm mode:
when alarm is set alarm relay is pulled. When
alarm is disabled alarm relay releases.
Latch mode:
when alarm mode is set alarm relay is pulled.
Alarm relay remains pulled even the alarm is disabled.
If latch
mode iset in models with alarm function, alarm should be
certainly disabled by pressing the set key.
Momentary mode
: In this mode, as an extra
the time seen in the adjacent box is displayed .
It may bet set between 300 msn and 5000msn.
When alarm is set, alarm relay remains pulled
until the end of this time then it is released.
This function may be used when continuous
alarm is not intended.