KAEL DC-MULTISER-1UH-DIN Manual Download Page 13

Te

chnical Information

1.

Measurements

Paramet

er

Measurement Range

Description

Voltage

Accuracy

0,5 %

0 ~  (±) 5 V

0,1 %

0 ~  (±)  60 V

0,1 %

0 ~  (±) 300 V

Current

0,1 %

0 ~  (±) 50.000 A

0,5 %

0 ~  (±) 600 V

0,5 %

0 ~  (±) 1000 V

For 

50 mV 

and

 100mV 

Shunt Output

Power

0,5 %

0 ~  (±) 60.000 kW

Ener

gy

0,5 %

0 ~  (±) 99.999.999,99 kWh

2.

Measurement Inputs

Paramet

er

Description

DC 

Voltage

300 KΩ

DC-MULTISER-1UH-DIN , DC-MULTISER-1UL-DIN , DC-MULTISER-2UH-DIN , DC-MULTISER-2UL-DIN  

2,5 MΩ

15 KΩ

May be used with shunt

Can be progrramed between 

0 ~  (±) 50.000 A/ 50 – 100mV  

DC 

Current

Input impedence

Operation Voltage

: 5 – 24Vdc  max. 30Vdc

3. Di

g

ital 

Input

 (1 

pc

)

4. R

elay Output

 (1 

pc

)

Operation Voltage

Max 

Load Voltage

Max 

Load Current

: 1NO 

and

  max.5A/250 

Vac

 

:  250 Vac/30Vdc

:  5A

5. Analog

ue Output

 (1 

pc

)

Voltage Output Type

 

Current Output Type

Accuracy

Load Capacity

  

: 0-10V, 2-10V, 0-5V, 1-5V

: 0-20mA, 4-20mA

: 0.2%

In current type

, max. 

load resistance

 

750 Ohm

 

In voltage type

load current

 max. 20 

mA

Updating Duration

:

 

200ms

6. RS485

T

ype

Proto

c

ol

Baud rate

Stop Bits 

Parity

: Half dublex

: Modbus RTU

: 2400,4800,9600,19200,28800,38400,57600 

or

 115200 : 

(0.5) , (1) , (1.5) 

or

 (2) 

: no , even , odd

Device

 No 

: 1 .....255

7.

Operation Voltage

(

DC-MULTISER-1UH-DIN, DC-MULTISER-1OH-DIN, DC-MULTISER-2UH-DIN, DC-MULTISER-2OH-DIN) 

(

DC-MULTISER-1UL-DIN, DC-MULTISER-1OL-DIN, DC-MULTISER-2UL-DIN, DC-MULTISER-2OL-DIN) 

Operation Voltage

(Un): 85 – 265  Vac   

or

 100 – 300 Vdc  

Operation Voltage 

(Un): 18 – 75  Vdc       

Operation Frequency

: 50/60 Hz

DC-MULTISER-1OH-DIN , DC-MULTISER-1OL-DIN , DC-MULTISER-2OH-DIN , DC-MULTISER-2OL-DIN  

Device Protection Class

  : IP 20

 

Terminal protection class

  : IP 00

Ambient Temperature

  : - 5 °C .... + 50 °C

  : 

to automat rail

Conneciton Type 

Dimensions

 

  : 105x90x59 mm

8.

Environment

105 mm

59 mm

34 mm

90

 m

m

Summary of Contents for DC-MULTISER-1UH-DIN

Page 1: ...1 1 1 2 1 1 1 2 1 1 1 2 1 1 DC MULTISER 2UH DIN DC MULTISER 2OH DIN DC MULTISER 2OL DIN DC MULTISER 2UL DIN Green Power Measurement Device Dimensions 105x90x59 mm Easy use via menus with shortcuts DC...

Page 2: ...f the system presents power quality problems insulation transformer or EMC filter should be used for auxiliary power supply Switch board cover or protection sheet With our DC MULTISER devices dc curre...

Page 3: ...t Connections AC DC AC DC Power Supply Relay Power Supply Digital Input Connections max 30Vdc Com DI2 Connect 1 5 2 5 mm cable Power Supply Optical Insulation Optik izolasyon DI1 Optical insulation An...

Page 4: ...eters menu At this stage HHHH is displayed in the line at the bottom of the screen while the password screen flashes When SET key is pressed 0000 figures are displayed in the most bottom line and the...

Page 5: ...arity no even odd Stop Bits May be set as 1 1 5 or 2 Baud rate It may be set as 2400 4800 9600 19200 28800 38400 57600 or 115200 Device No 001 255 SET SET SET SET Relay Contact Position nO normally sw...

Page 6: ...Momentary mode In this mode as an extra the time seen in the adjacent box is displayed It may bet set between 300 msn and 5000msn When alarm is set alarm relay remains pulled until the end of this tim...

Page 7: ...e Curr Over dELy It may be set between 1 300 sec If the measured voltage exceeds the set voltage value the relay contact is released at the end of the delay time Over Current Mistake Repair Time Curr...

Page 8: ...rrent Delay Time Curr Undr dELy It may be set between 1 300 sec If the measured current falls below set current value relay contact is released at the end of the delay time Low Current Mistake Repair...

Page 9: ...s disabled alarm relay releases Latch mode when alarm mode is set alarm relay is pulled Alarm relay remains pulled even the alarm is disabled If latch mode iset in models with alarm function alarm sho...

Page 10: ...disregarding the symbol of the measured value In this mode measred value may be set between 100 This menu is displayed when PErC is selected in the Analogue measurement range An rAng This screen does...

Page 11: ...utputs Measurement Value Analogue output value 2 5 V 5 V 5 V Voltage V 0 10 V Example 1 Voltage Full Scale Setting Analogue Parameters Voltage Output Type Analogue Operation range setting PERC PERC up...

Page 12: ...arm mode 10 sec 10 sec Over Current Set Value Curr Over Val Over Current Alarm Permit Curr Over EnbL Over Current Alarm Mode Setting Curr Over tYPE Over Current Delay Time Curr Over dELy Over Current...

Page 13: ...t Type Current Output Type Accuracy Load Capacity 0 10V 2 10V 0 5V 1 5V 0 20mA 4 20mA 0 2 In current type max load resistance 750 Ohm In voltage type load current max 20 mA Updating Duration 200ms 6 R...

Reviews: