Calibration
37
Figure A-2: Auto-Calibration Flow
Set analog front end input
to onboard calibration source
Set calibration source
to ground
Capture data and calculate
offset compensation parameters
Are all channels and
all ranges complete?
No
Set calibration source
to calibration voltage
Capture data and calculate
gain compensation parameters
Yes
Are all channels and
all ranges complete?
No
Yes
Set analog front end input
to SMB connector
Auto-calibration start
Auto-calibration
complete
Summary of Contents for PCIe-69834
Page 8: ...viii List of Figures This page intentionally left blank...
Page 10: ...x List of Tables This page intentionally left blank...
Page 18: ...8 Introduction Figure 1 3 PCIe 69834 I O Array All I O connectors are SMB snap on...
Page 20: ...10 Introduction This page intentionally left blank...
Page 44: ...34 Operations This page intentionally left blank...
Page 48: ...38 Calibration This page intentionally left blank...