12 device description
12.1 Technical data
12.1.1 Inputs
1
Selectable in
°F.
2
Depending on supporting points.
12.1.2 Temperature compensation
12.1.3 Measuring circuit monitoring
12.1.4 2-electrode systems
Main input
Indication range
Accuracy
Temperature error
µS/cm
0.000 — 9.999
00.00 — 99.99
000.0 — 999.9
0000 — 9999
≤
0.6% of range + 0.3 µS x cell
constant (K)
0.2%/10°C
mS/cm
0.000 — 9.999
00.00 — 99.99
000.0 — 999.9
0000 — 9999
≤
0.6% of range + 0.3 µS x cell
constant (K)
0.2%/10°C
k
Ω
x cm
0.000 — 9.999
00.00 — 99.99
000.0 — 999.9
0000 — 9999
≤
0.6% of range + 0.3 µS x cell
constant (K)
0.2%/10°C
M
Ω
x cm
0.000 — 9.999
00.00 — 99.99
000.0 — 999.9
0000 — 9999
≤
0.6% of range + 0.3 µS x cell
constant (K)
0.2%/10°C
Secondary input
Measuring range
Accuracy
Temperature error
Temperature Pt100
(automatic detection)
-50 to 250°C
1
±
0,5_K (up to 100 °C)
±
0,8_K (as of 100 °C)
0.05%/10°C
Temperature Pt1000
(automatic detection)
±
0,5_K (up to 100 °C)
±
1,0_K (as of 100 °C)
Temperature
NTC/PTC
max. 4 kOhm
Input via table with 20 value
pairs, through setup program
≤
0.3%
2
0.05%/10°C
Type of compensation
Range
Linear 0 — 8%/°C
-10 to 160°C
ASTM D1125 - 95 (highly-purified water)
0 to 100°C
Natural water (EN 27 888)
0 to 36°C
Reference temperature
adjustable from 15 to 30°C; preset to 25°C (standard)
Inputs
Over/underrange
Short-circuit
Cable break
Conductivity
yes
depending on range
depending on range
Temperature
yes
yes
yes
Cell constant
[1/cm]
Setting range of
relative cell constant
Resulting usable
range [1/cm]
0.01
20 — 500%
0.002 — 0.05
0.1
0.02 — 0.5
1.0
0.2 — 5
3.0
0.6 — 15
10.0
2.0 — 50
Summary of Contents for AQUIS 500 CR