7 Commissioning
60
Concluding device settings
Input for
conductivity
Cell constant:
0.01
Relative cell constant:
102.5%
Offset for the meas. range:
0.0°C
Temperature compensation: ASTM D1125-95
Filter time constant:
As required
Calibration interval:
As required
Input for
temperature
Sensor type:
Pt100/Pt1000
Unit:
°C
Filter time constant:
00:00:02
Offset:
0.0°C
Analog output 1
Signal selector:
Main value
Signal type:
4 — 20 mA
Scaling start:
0 MOhm*cm
Scaling end:
20 MOhm*cm
During calibration:
As required
In event of error:
As required
in Hold mode:
As required
Safe value:
As required
Simulation:
As required
Simulation value:
As required