6 Operation
44
6.11 Controller functions
Simple
switching
functions
In the JUMO AQUIS 500, simple switching functions, such as alarm contacts
and limit comparators or the signal from the calibration timer, are configured at
the parameter level, through the parameters for “Switching output 1 or 2”.
The parameters for the controller channel 1 or 2 respectively must then be set
to “Off”.
Higher-level
control
functions
Higher-level control functions are configured at the parameter level, through
the parameters for “Controller channel 1 or 2”.
The controller channel parameter must then be set to “Controller 1 or 2”.
Operator level
parameters
Switching output 1 / 2
Explanation
none
no switching function and
no control function required
Controller 1
the device should have “higher-level” control
Controller 2
the device should have “higher-level” control
Controller alarm 1 / 2
Controller alarm
main variable
main variable
main variable
main variable
Temperature
Temperature
Temperature
Temperature
Sensor fault
Calibration timer
Autorange
USP
USP pre-alarm
PH.-EUR
PH.-EUR pre-alarm
“simple” switching functions
Controller channel 1 / 2
Limit value
Pulse width
Pulse frequency
Continuous
Modulating
“higher-level” control functions
Off
must be selected if “simple” switching
functions are required