75
11 Appendix
11.1 Operator level parameters
If a number of device parameters have to be modified in the device, then it is
advisable to note them in the table below, and then modify these parameters
in the sequence given.
The following list shows the maximum number of parameters that
can be altered. Depending on the configuration, some of the
parameters will not be visible, i.e. not alterable (editable) for your
device.
Parameter
Selection / value range
Factory setting
New
setting
Conductivity input
Cell constant
0.01 / 0.1 / 0.5 / 1.0 / 3.0 / 10.0
1
/
cm
Relative cell constant
20.0 — 100.0 — 500%
Offset for the meas. range -200 to 0 to +200
Temperature
compensation
LINEAR
NATURAL WATER (permissible temperature
range 0 to 36°C
as per EN 27 888)
ASTM D1125-95
Temperature coefficient
0.00 — 2.20 — 8.00%/°C
Reference temperature
15.0 to 25.0 to 30°C
Filter time constant
0 — 2— 25 sec
Calibration interval
0— 999 days (0 = switched off)
Temperature input
Sensor type
Pt100/Pt1000
CUSTOMIZED
NO SENSOR
Unit
°C
°F
Filter time constant
00:00:00 — 00:00:02— 00:00:25 H:M:S
Offset
-20.0 to 0.0 to +20.0°C
Binary input
Function
NO FUNCTION
KEY LOCK
HOLD MODE
ALARM STOP (controller)
Controller channel 1
Controller type
LIMIT
PULSE WIDTH
PULSE FREQ.
CONTINUOUS
MODULATING
OFF
Setpoint
0 — 9999
MIN / MAX contact
(increasing / decreasing
characteristic)
MIN CONTACT
MAX CONTACT