4 Calibrating a pH combination electrode
20
4.5
Display of the electrode state
✱
Briefly press the
key.
The state of the pH combination electrode will be displayed for approx. 3
seconds.
4.6
Parameters in the combination electrode display
Show
asymmetry
✱
Briefly press the
key.
The state of the pH combination electrode will be displayed for approx. 3
seconds after every calibration.
100%
The combination electrode is in very good condition
90 — 30%
The state of the combination electrode is satisfactory
less than
30%
The combination electrode is severely aged or contaminated.
If there is no improvement after cleaning (see Chapter 7.2
“Cleaning”, page 27) and a subsequent calibration, then the
combination electrode must be replaced.
2
CAL
3
pH
mV
mV
H
rH
ATC
ok
CAL
AL Logg
pH
mbar
The evaluation takes into account both the asymmetry and the slope of the
combination electrode.
Permissible
range
±60 mV
Optimum
0 mV
2
CAL
3
pH
mV
mV
H
rH
ATC
ok
CAL
AL Logg
pH
mbar
Asymmetry -30 mV
Contamination can seriously worsen the asymmetry of the combination
electrode.