Operation
83
9. When
[Enter temperature]
is pressed, the dialog window for entering
the temperature value and correction value opens.
10. Enter the required value and confirm your entry.
The correction value is active immediately.
The internal temperature sensor is adjusted with a correction point.
8.12.2 Adjust the internal temperature sensor - multi-point calibration
This section describes how to adjust the internal temperature sensor of the
device with multiple correction points. The adjustment can be performed with
up to ten correction points.
The device is switched on.
The bath tank is at least two thirds full.
A calibrated reference thermometer is available.
1. Initially proceed in the same way as for the 1-point adjustment.
2. Go to the Home screen and enter the setpoint temperature for the next
correction point.
3. Wait until the temperature in the bath has stabilized and is shown on the
display of the circulator.
4. Go to the ATC submenu
<ATC internal>
and press the
[Correction point 1]
button or the button of the last correction point that
was added.
The correction point opens.
5. Press the button for a new correction point and proceed in the same way as
described for the 1-point adjustment.
6. Repeat steps 2 to 5 for each additional correction point and confirm your
input in each case.
The correction values are active immediately.
Up to ten correction points can be created for the adjustment.
The multi-point adjustment is completed. The internal temperature sensor is
adjusted.