Operation
75
8.9 Working with the programmer
Temperature control profiles are defined temperature profiles with a setpoint
temperature as well as a predefined temperature control duration or a
predefined gradient. The programmer calculates a temperature ramp from these
programmed values.
The programmer can start these temperature control profiles directly or with a
time delay. The number of repetitions can be set, as can the start time and the
days of the week on which the temperature control profile should run.
In order for programmed temperature control profiles to run automatically, the
device must remain switched on.
NOTE
Falling bath fluid level!
In the case of a prolonged temperature control application, the level of bath
fluid in the bath tank can fall below the alarm limit due to evaporation. A low
liquid level alarm is triggered and temperature control application stops.
• In the event of a prolonged temperature control application, regularly
check the fill level of bath fluid in the bath tank
• If the level is low, refill the bath fluid
8.9.1 Setting the timer
The timer can be used to program the duration of a tempering from 0 to 999
minutes and to program the start time. The target temperature is set as the
setpoint temperature.
After the set duration has elapsed, the device switches to the previously defined
state:
• Standby mode
• Maintain target temperature
• Temper to original setpoint temperature
The device is switched on.
1. Call up the
<Main menu>
.
2. Call up the submenu
<Use programmer>
then the menu option
<Timer>
.
The timer setting options are displayed in the submenu.
No parameters can be changed when the timer is active. First deactivate the
timer.
3. Use the
[Plan start]
button to set the desired start time and date then
confirm your input.