Initial operation
58
In order to gently control the temperature of a sample in an external system, or
to protect a glass reactor from thermal stress, for example, maximum
permissible temperature differences can be specified for the heating phase and
the cooling phase.
7.9.5 Setting band limit
The upper and lower band limit is defined in °K. In the heating phase the set
value will be added to the external actual value. In the cooling phase the set
value will be subtracted from the external actual value. The setting range is
between 1 and 200 °K.
The device is switched on.
1. Call up the
<Main menu>
.
2. Call up the submenu
<Adjust safety>
.
3. Set the upper and lower band limit.
The band limits are set and immediately active.
7.10 Connect external temperature sensor
A Pt100 temperature sensor can be attached to the rear of the device.
The device is switched off.
A Pt100 temperature sensor with suitable connection is available.
1. Connect the Pt100 to the ext.Pt100 socket on the back of the device.
2. Switch the device on.
3. Call up the submenu
<Determine thermodynamics>
and the menu
option
<Adjust controller>
.
4. In the menu option
<Control>
activate the interface
[External control]
.
5. Call up the menu option
<External>
.
6. Set the control parameter for external control.
The external temperature sensor is connected and ready for operation.
B