Appendix
102
Error code Description
Solution
-144
The set temperature limit for
subtemperature alarm was exceeded.
•
Decrease temperature limit of
subtemperature alarm or increase
temperature setpoint.
-503
External setpoint value is set via
EPROG, but no analog module is
connected.
•
Connect analog module or deactivate
external setpoint specification using
EPROG.
-504
External actuating variable is set via
EPROG, but no analog module is
connected.
•
Connect analog module or deactivate
external actuating variable using
EPROG.
-505
The analog module transmits an
invalid setpoint.
•
Check EPROG settings.
-1109
Viscosity of the bath fluid too high or
the circulation rate too low.
•
Check the bath fluid for its suitability
in the temperature range being used.
-1305
The rotational speed limit for the
heating capacity block is not reached.
Motor defective or fluid viscosity too
high.
•
Check pump setting and adjust if
necessary.
•
Check viscosity of the bath fluid and
adjust if necessary.
-1431
Compressor current too low.
•
Check the mains voltage and align it
with the nominal voltage of the
device.
•
Check the mains cable of the
refrigeration unit for damage and
replace if necessary.
•
Check ambient temperature and
reduce if necessary.
•
Check CAN bus cable for damage and
replace if necessary.
•
The specified voltage tolerance of the
device must not be exceeded. If
necessary Adjust the mains voltage to
the nominal voltage of the device.
•
If the fault has not been remedied,
contact Technical Service.
-1501
Timeout on serial interface.
•
If the watchdog is activated, check
whether the temperature setpoint is
sent to the device cyclically at least
every 30 seconds.
-2426
Evaporation temperature ha fallen
below the warning threshold.
•
Check pump setting