Appendix
71
-2426
Evaporation temperature has fallen
below the warning threshold.
•
Check pump setting
•
Check hose cross-section of connection
to application
•
Check that connection to application is
clear
•
Check suitability of bath fluid
11.3 Error messages in configuration process
If errors occur during a configuration process or during a firmware update, these
are shown on the display as error codes in ticker text.
Error code Description
Solution
CFG error
Error during configuration.
•
Repeat the process.
•
Replace USB stick if this occurs again.
•
If the fault has not been remedied,
contact Technical Service.
ProG error
Error during firmware update.
•
Repeat the process.
•
Replace USB stick if this occurs again.
•
If the fault has not been remedied,
contact Technical Service.