![Julabo CORIO CP-1000F Original Operating Manual Download Page 32](http://html1.mh-extra.com/html/julabo/corio-cp-1000f/corio-cp-1000f_original-operating-manual_2030700032.webp)
Product description
32
5.8.3 Bath fluids
The most important criterion when selecting the bath fluid is the working
temperature range in which the application is operated.
• Selection of the bath fluid must ensure that the flash point is never
exceeded when it comes into contact with the ambient air.
• Recommended bath fluids and further information can be found on our
website
NOTE
No liability accepted for usage of bath fluids that are not
suitable!
Unsuitable bath fluids that are not approved by JULABO can damage the
water bath.
• Use bath fluids that are recommended by JULABO
• Before filling, check the parts that are in contact with the medium for
compatibility with the bath fluid
• Do not exceed the maximum permissible viscosity during operation
• Consult JULABO before using a bath fluid other than the recommended
one
NOTE
Material damage caused by freezing water!
Freezing water can damage the device and the sample.
• When using water as the tempering fluid, do not temper below 5°C
Water as bath fluid
• Water can be used for working temperatures from +5 °C to +90 °C
• Recommended: Use only ultrapure or distilled water.
• When using ultrapure or distilled water, add 0.1 g Na
2
CO
3
per liter of water.
Recommended bath fluid
• JULABO Thermal G