28
Maintenance
4. Cuff Cleaning and Disinfection:
A) Spread the cuff (skin-contact surface) upwards onto a clean table. Use a damp clean cloth
(water-based) to wipe the skin-contact surface with a force.
B) Soak the cloth clean with drinking water and wring it dry. Repeat A) with the damp cloth
(water-based) for 3 times.
C) Apply 70%-80% alcohol to a new cloth (or 75% alcohol cotton-ball), use it to
wipe the skin-contact surface with a force. Then soak the cloth with the alcohol again
(or change a new 75% alcohol cotton-ball), repeat the disinfection procedure for 3 times.
D
)
When
the
disinfection
towards
the
skin
-
contactsurface is
finished
,
wipe
thenon
-
skin
contact surface
with
a
cloth (alcohol-based) or alcohol cotton-ball
thoroughly
for
3 times.
E
)
Leave
the
cuff
naturally
dry
,
then
it
is
ready
for
reuse
.
Notice: Do not soak in water or splash water on it.
5. Do not use petrol, thinners or similar solvents.
6. Remove batteries when not in operation for an extended period of time.
7. Do not disassemble product.
8. It is recommended the performance should be checked every 2 years.
9. Expected service life: Approximately three years at 10 tests per day.
10.No service and maintenance while it is in use and maintenance only be performed by service
personnel. Service and maintenance require parts, repair, technical support will be provided.
Maintenance
27
1. Avoid dropping, slamming, or throwing the unit.
2. Avoid extreme temperatures. Do not expose unit directly under sunshine.
3. When cleaning the unit, use a soft fabric and lightly wipe with mild detergent.
Use a damp cloth to remove dirt and excess detergent.
30
29
Specifications
Specifications
Product Description
Model
Display
Measurement Method
Pressurization
Memory
Function
Power Source
Battery Life
Unit Weight
Unit Dimensions
Cuff Circumference
Oscillometric Method
Arm-type Fully Automatic Blood Pressure Monitor
DBP-6277B
LCD Digital Display Size:83.1mm
×
53.1mm (3.27" x 2.09")
2x60 Memories in Two Groups with Date and Time
Irregular Heartbeat Detection
WHO Classification Indicator
Low Battery Detection
Automatic Power-Off
Voice
3 AAA batteries or Medical AC Aadapter (DC5.0V, 1000mA)
(recommended, not provided)
Approximately 2 months at 3 tests per day
Approx.213g (7.51 oz.) (excluding battery)
Approx.149.8 x 80.4 x 45.7mm (5.90
"
x 3.17
"
x 1.80
"
)(L x W x H)
Approx.135 (W) x 485(L) mm
(Medium cuff: Fits arm circumference 22-36 cm)
Operating Environment
Humidity
Temperature
1
~ 40
(50
~104
)
0
℃
℃
℉
℉
15% ~ 93%RH
Pressure
800hPa~1060hPa
Automatic Pressurization
Measurement Range
Pulse
Systolic Pressure
Pulse
60mmHg
~
2
60
mmHg
30 ~ 180 Beats/Minute
±
5%
Diastolic Pressure
30mmHg
~
20
0
mmHg
Pressure
0mmHg
~
299mmHg
Pressure
±
3mmHg
Backlight
Last 3 Tests Average
Bluetooth
Classification
Ingress Protection
rating
Battery Shelf life
Battery Storage
Temperature
Internal Powered Equipment,Type BF , Cuff is the Applied Part
IP 20,Indoor Used Only
60 months
-25
℃
~55
℃
(-13
℉
~131
℉
)
Storage Environment
Humidity
Temperature
-25
℃
~55
℃
(-13
℉
~131
℉
)
≤
93% RH
Transport Environment
Humidity
Temperature
-25
℃
~55
℃
(-13
℉
~131
℉
)
≤
93% RH
Bluetooth
Version
Modulation Type
GFSK
5.0.1 BT Signal mode
Operation frequency
2
.
4GHz
(
2400
~
2483
.
5MHz
)
Antenna gain
0.5 dBi
This Blood Pressure Monitor complies with the European regulations and bears the
CE mark“CE 0197”.This blood pressure monitor also complies with mainly
following standards
(included but not limited):
Safety standard:
EN 60601-1 Medical electrical equipment part 1: General requirements for safety
EMC standard:
EN 60601-1-2 Medical Electrical Equipment -- Part 1-2: General Requirements For Basic
Safety And Essential Performance -- Collateral Standard: Electromagnetic Disturbances –
Requirements And Tests.
Performance standards:
IEC80601-2-30, Medical electrical equipment – Part 2-30: Particular requirements for the
basic safety and essential performance of automated non-invasive sphygmomanometers.
EN 1060-3 Non-invasive sphygmomanometers - Supplementary requirements for
electromechanical blood pressure measuring systems.
ISO 81060-2, non-invasive sphygmomanometers - part 2: clinical validation of automated
measurement type.
Specifications are subject to change without notice.
Bandwidth
2.0 MHz