Unit Operation
Unit Operation
Unit Operation
Low Battery Indicator
The unit will broadcast "Low Battery" when battery life is depleting and unable to inflate cuff for
testing. The " " appears simultaneously for approximately 5 seconds prior to shutting
off. Replace batteries at this time. No memory loss will occur throughout this process.
Memory Deletion
Memory for a selected group may be deleted while in Memory Check mode. Press and hold
the " SET " button for approximately 3 seconds to delete all memory records from the
selected group. with voice broadcast "Memory Clear" and then transfer into
testing mode. Press the "START/STOP " button to turn the unit off.
Note: Memory cannot be recovered once it has been deleted.
Unit Operation
19
In the power down state, press and hold the " START/STOP" button, and theninstall
the batteries. Until the LCD screen is full, release the " START/STOP" button.
When the LCD screen displays the double zero, the bloodpressure meter is in static state.
Software version is displayed:10 is a software version in the figure.
Static Pressure Measurement
Note: Only Service personnel permitted to access to this mode, the mode unavailable in
normal use.
Bluetooth requirements
The monitor requires a device with:
. Bluetooth 4.0 or later
. Android 5.0 or later
. IOS 9.0 or later
And works with:
. iphone , iPod, iPad
. Android Phones and Tablets
1. Open the “blood pressure monitor” and follow the pairing instructions shown on your smart
phone.
The date and time on your monitor will automatically be set when you pair it with your
smart device.
2. Confirm that your monitor is connected successfully.
When your monitor is connected successfully to your smart phone, it will be display like
below.
3. Press the BP [START/STOP] button to turn your monitor off.
-Pairing
your monitor with a
S
mart Device
Wireless communication
Frequency range : 2.4 Ghz (2400-2483.5 MHz)
Modulation : GFSK
Antenna gain:0.5dBi
Troubleshooting
Problem
Solution
" "displayed
Connection
failure./
Data is not being
transmitted
Cuff is too tight or not
properly positioned on
the arm
Firmly reposition cuff
approximately
(
)
above the elbow joint
( See Page 12)
1-2cm 1/2"
Inaccurate test results due to
body movement or monitor
movement
Sit in a relaxed position with
arm placed near heart. Avoid
speaking or moving body parts
while testing. Make sure the
monitor unit is placed in a
stationary position throughout
the testing period.
( See Page 7)
Blood pressure
results are not
within typical range
Cuff fails to inflate properly
The blood pressure monitor
might not be porperly placed
within the smart device’s
tranmission range and is too far
from the smart device.
Make sure hose is properly
fastened to cuff and monitor unit
If there are no causes of data
transmission interference found
near the blood pressure monitor,
move the blood pressure monitor
within 16ft.(5m) of the smart
device and try again
Possible Cause
Pressurization is over
cuff rated pressure
300mmHg
The blood pressure did not pair
successfully to the smart device
Read user manual carefully and
re-test properly.
Try to pair the devices once again.
Improper operation
The application on the smart
device is not ready.
Read user manual carefully and
re-test properly.
Check the application then try
sending the data again.
1. Download the free “JoyHealth”
App: On your mobile phone or table
go to www.sejoy.com.
2.Open the App on your phone or tablet. If requested, you should enable Bluetooth on your
device. You can enable Bluetooth under the Settings menu on your smart phone or table.
3. Create a new user login, or login with your existing user name and password.
4. Selection device “Blood pressure monitor”.
Bluetooth connection
-Using
for
the
first
time
-Transfer your readings
1. As soon as your measurement is complete, open the app on your smart phone to transfer
your reedings.
Notr: On the paired smartphone, Bluetooth must be enabled.
2. You can view your blood pressure readings on the app.
AM
BP
AM
BP
mmHg
Bluetooth
SMART