JOYTECH DBP-6277B Owner'S Manual Download Page 6

21

20

22

Unit Operation 

Unit Operation 

Unit Operation 

Low Battery Indicator

The unit will broadcast "Low Battery" when battery life is depleting and unable to inflate cuff for 
testing.  The "             " appears simultaneously for approximately 5 seconds prior to shutting 
off.  Replace batteries at this time.  No memory loss will occur throughout this process.

Memory Deletion 

Memory for a selected group may be deleted while in Memory Check mode.  Press and hold 

the " SET " button for approximately 3 seconds to delete all memory records from the 

selected group.  with voice broadcast "Memory Clear" and then transfer into 

testing mode.  Press  the "START/STOP " button to turn the unit off.  

Note:  Memory cannot be recovered once it has been deleted.

Unit Operation 

19

In the power down state, press and hold the " START/STOP" button,  and theninstall 

the batteries.  Until the LCD screen is full, release the " START/STOP" button. 

When the LCD screen displays the  double zero, the bloodpressure meter is in static state. 

Software version is displayed:10 is a software version in the figure.

Static Pressure Measurement

Note: Only Service personnel permitted to access to this mode, the  mode unavailable in 
            normal use.

Bluetooth requirements

   The monitor requires a device with:

   . Bluetooth 4.0 or later

   . Android 5.0 or later

   . IOS 9.0 or later 

   And works with:

   . iphone , iPod, iPad

   . Android Phones and Tablets

1. Open the “blood pressure monitor” and follow the pairing instructions shown on your smart

    phone.

   The date and time on your monitor will automatically be set when you pair it with your 

   smart device.

2. Confirm that your monitor is connected successfully.

   When your monitor is connected successfully to your smart phone,  it will be display like 

   below.

3. Press the BP [START/STOP] button to turn your monitor off.

   

-Pairing

 your monitor with a 

S

mart Device

            Wireless communication

             Frequency range : 2.4 Ghz (2400-2483.5 MHz)

             Modulation : GFSK

             Antenna gain:0.5dBi

 

Troubleshooting

Problem

Solution

"        "displayed

Connection

 

failure./

Data is not being 
transmitted

Cuff is too tight or not 
properly positioned on 
the arm

Firmly reposition cuff 
approximately

 (

above the elbow joint
( See Page 12)

1-2cm 1/2"

Inaccurate test results due to 
body movement or monitor 
movement

Sit in a relaxed position with 
arm placed near heart.  Avoid 
speaking or moving body parts 
while testing.  Make sure the 
monitor unit is placed in a 
stationary position throughout 
the testing period.
( See Page 7)

Blood pressure 
results are not 
within typical range

Cuff fails to inflate properly

The blood pressure monitor 
might not be porperly placed 
within the smart device’s 
tranmission range and is too far
 from the smart device.

Make sure hose is properly 
fastened to cuff and monitor unit

If there are no causes of data
transmission interference found 
near the blood pressure monitor,
move the blood pressure monitor
within 16ft.(5m) of the smart 
device and try again

Possible Cause

Pressurization is over  
cuff rated pressure 
300mmHg 

The blood pressure did not pair 
successfully to the smart device

Read user manual carefully and 
re-test properly.

Try to pair the devices once again.

 Improper operation

The application  on the smart 
device is not ready. 

Read user manual carefully and 
re-test properly.

Check the application then try 
sending the data again.

1. Download the free “JoyHealth”

 

App: On your mobile phone or table 

go to www.sejoy.com.

2.Open the App on your phone or tablet. If requested, you should enable Bluetooth on your

 device. You can enable Bluetooth under the Settings menu on your smart phone or table.

3. Create a new user login, or login with your existing user name and password.  

4. Selection device  “Blood pressure monitor”.

Bluetooth connection

-Using

 

for

 

the

 

first

 

time

-Transfer your readings

1. As soon as your measurement is complete, open the app on your smart phone to transfer

   your reedings.

Notr: On the paired smartphone, Bluetooth must be enabled.

2. You can view your blood pressure readings on the app.

   

AM

BP

AM

BP

mmHg 

Bluetooth

SMART

Summary of Contents for DBP-6277B

Page 1: ...le Operation 1 Do not drop the unit Protect it from sudden jars or shocks 2 Do not insert foreign objects into any openings 3 Do not attempt to disassemble the unit 4 Do not crush the pressure cuff 5...

Page 2: ...ambulance or helicopter 30 Contains small parts that may cause a chocking hazard if swallowed by infants 31 Please align the polarities of each battery with the ve and ve signs imprinted on the batte...

Page 3: ...everal minutes prior to testing Sit down in a quietplace comfortably back and arm support on a desk or table withyour legs uncrossed your arm resting on a firm and your feet flat on the floor See Figu...

Page 4: ...es it will automatically save all information and shut off Unit Operation System Settings With power off press SET button to activate System Settings The Memory Group icon flashes 1 Select Memory Grou...

Page 5: ...memory can be viewed by pressing and holding the MEM button Upon activating test results you can press the MEM button to scroll through all test results stored in memory With power off press the MEM...

Page 6: ...GFSK Antenna gain 0 5dBi Troubleshooting Problem Solution displayed Connection failure Data is not being transmitted Cuff is too tight or not properly positioned on the arm Firmly reposition cuff app...

Page 7: ...Many factors including diet stress cuff placement etc may affect an individual s blood pressure Q Should I apply the cuff to the left or right arm What is the difference A Either arm can be used when...

Page 8: ...ate and Time Irregular Heartbeat Detection WHO Classification Indicator Low Battery Detection Automatic Power Off Voice 3 AAA batteries or Medical AC Aadapter DC5 0V 1000mA recommended not provided Ap...

Page 9: ...y 50 60 Hz magnetic field IEC 61000 4 8 30 A m 50Hz or 60Hz Power frequency magnetic fields should be at levels charactertic of a typical location in a typical comme rcial or hospital environment 30 A...

Page 10: ...advice Pressure calibration will be carried out when this product leaves the factory Patients can use the method described in the section Verify Manometer Pressure Accuracy to verify the accuracy If...