Automatic set
Auto set is one of the advantages digital oscilloscopes have. When you push the Auto
button, the oscilloscope will identify the type of waveform (sine or square wave) and
adjust controls according to input signals so that it can accurately display the wave-
form of the input signal.
The Auto function examines all
channels for signals and displays
corresponding waveforms.
Auto determines the trigger
source according to the following
conditions :
If multiply channels get signals,
the oscilloscope will use the
channel with the lowest frequen-
cy signal as the trigger source.
If no signals are found, the oscil-
loscope will use the lowest
-
numbered channel displayed in Auto Scale as the trigger source.
If no signals are found and no channels are displayed, the oscilloscope will display
and use Channel 1 as the trigger source.
•
Default settings
Functions
Settings
Cursor
Off
Display Format
Set to YT
Horizontal Position
Adjusted
SEC/DIV
Adjusted
Trigger Level
Set to 50%
Trigger Mode
Auto
Trigger Source
Adjusted
Trigger Slope
Adjusted
Trigger Type
Edge
Vertical Bandwidth
Full
Vertical Coupling
Unchanged
VOLTS/DIV
Adjusted
Menu or Sys-
Option, Button or
Default Setting
Cursor
Type
Off
Source
CH1
Horizontal (amplitude)
±4div
Vertical (time)
±4div
Display
Format
YT
Horizontal
Position
0.00s
SEC/DIV
500μs
Measure
On or Off
Off
Trigger (Edge)
Source
CH1
Slope
Rising
Mode
Auto
Level
0.00v
Vertical Sys-
tem,
All Channels
Bandwidth Limit
Unlimited
Coupling
AC
Probe Attenuation
1X
Position
0.00div (0.00V)
VOLTS/DIV
1V