Joy-it JT-DMSO2D72 Manual Download Page 10

3. 

Press the 

Auto

 button and you should see within a few seconds a square wave 

of about 2V peak

-

to

-

peak at 1KHz in the display

. Repeat the steps to observe CH2.

 

• 

Probe check

 

Safety

 

When  using  the  probe,  keep  your  fingers  behind  the 

guard  on  the  probe  body  to  avoid  electric  shock.  Do 

not touch metallic portions of the probe head while it 

is connected to a voltage source. Connect the probe to 

the  oscilloscope  and  connect  the  ground  terminal  to 

ground before you start any measurements

Manual Probe Compensation

 

Upon the first connection of a probe and an input channel, you should manually 

perform this adjustment to match the probe to the input channel. Uncompensated 

or  wrong  compensated  probes  may  lead  to  errors  or  faults  in  measurement.  To 

adjust the probe compensation, follow the steps below.

 

1. Press 

Channel

 button to enter channel setting menu, Set the Probe option at-

tenuation in the channel menu to 10X. Set the switch on the probe to 10X and con-

nect the probe to Channel 1 on the oscilloscope. If you use the probe hook

-

tip,  the 

hook end should be removed, the probe pin should be inserted into the Gen Out 

output terminal, and the probe grounding clamp should be clamped on the metal 

outer ring of the Gen Out output terminal. The oscilloscope with signal generator 

function needs to set output signal as 2V @ 1KHz square wave; The Gen Out termi-

nal of the oscilloscope without signal generator function automatically outputs 2V 

@ 1KHz square wave. Press the 

Auto

 button.

 

2. Check the shape of the displayed waveform. 

 

 

 

 

 

 

 

 

 

3. If  necessary,  use  a  nonmetallic  screwdriver 

to  adjust  the  variable  capacity  of  your  probe 

until  the  shape  of  the  waveform  turns  to  be 

the same as the above figure. Repeat  this step 

as necessary. See the figure below for the way 

of adjustment.

 

Probe  Attenuation  Setting

Probes  are  of 

various  attenuation  factors  which  affect  the 

vertical  scale  of  the  signal.  The  Probe  Check 

function is used to verify if the Probe attenua-

tion option matches the attenuation of the probe.

 

Compensated correctly

 

 

Overcompensated

 

 

Undercompensated

 

Summary of Contents for JT-DMSO2D72

Page 1: ...cope is compact portable and flexible It has a TFT LCD color display with a resolution of 320 x 240 pixels and adjustable backlight brightness The maximum real time sampling rate can be up to 250 MSa...

Page 2: ...ards use this product only as specified Only qualified personnel should perform maintenance Use suitable power cord Use only the power cord specified for this prod uct and certified for the country of...

Page 3: ...pected failures If you suspect there is damage to this product have it inspected by qualified service personnel Assure good ventilation Do not operate in wet damp environments Do not operate in an exp...

Page 4: ...r GND Input impedance DC coupled 25pF 3pF 1M 2 Probe attenuation 1X 10X Supported probe attenuation factors 1X 10X 100X 1000X Input protection voltage 150 VRMS Measurement Values between cursors Volta...

Page 5: ...t 624 g Arbitrary Waveform Generator Waveform Frequency Sine 1Hz 25MHz Square 1Hz 10MHz Ramp 1Hz 1MHz Expon 1Hz 5MHz Sampling rate 250 MSa s Amplitude 2 5 Vpp 50 5Vpp High impedance Frequency resoluti...

Page 6: ...2 100uA 4 000A 1 8 2 1mA 10 00A 3 2 10mA Overload protection self restoring fuse 200mA 250V 4A and 10A range no fuse AC Current 40 00mA 1 3 2 10uA 400 0mA 1 8 2 100uA 4 000A 2 3 1mA 10 00A 3 5 10mA Fr...

Page 7: ...r receiving Check the shipping container for damage Keep the damaged shipping contain er or cushioning material until the contents of the shipment have been checked for completeness and the instrument...

Page 8: ...r easy operation and observation When the instrument is not in use the user can close the support foot to facilitate placement or han dling After adjusting the rack the instrument can be suspended on...

Page 9: ...r key backlight is red and flash if the battery is installed inside the oscilloscope and the battery power is not filled the power key backlight is red if the battery is installed inside the oscillosc...

Page 10: ...tch on the probe to 10X and con nect the probe to Channel 1 on the oscilloscope If you use the probe hook tip the hook end should be removed the probe pin should be inserted into the Gen Out output te...

Page 11: ...ntal and vertical scales of the oscilloscope automatically and sets the trigger coupling type position slope level and mode etc to acquire a stable waveform display Channel Channel setting menu Time H...

Page 12: ...signal genera tor function always outputs 2V 1KHz square wave Shortcut keys Long press this button to enter the menu and choose shortcut key function After setting press this button sin gle time to re...

Page 13: ...hannel displayed in Auto Scale as the trigger source If no signals are found and no channels are displayed the oscilloscope will display and use Channel 1 as the trigger source Default settings Functi...

Page 14: ...or Left buttons to move the waveform right or left The key resolution vary according to time base Press AUTO key can make the horizontal position return to zero 3 Mode Y T X Y Roll Scan Y T YT format...

Page 15: ...er Edge trigger distinguishes the trigger points by seeking the specified edge rising falling rising falling and trigger level Press Trig button to enter trigger menu Trigger Source Select the trigger...

Page 16: ...el It sets the amplitude level the signal must cross to cause an acquisiti on when using the Edge or Pulse Width trigger Press Trig button to enter and press Up or Down direction keys to change the tr...

Page 17: ...mV division 300mV Cursor measurement The cursor measurement has two parallel lines on the screen and move two lines to measure the time and voltage parameters of the input signal The result of cursor...

Page 18: ...er of the screen Attention In the measurement result the yellow font is the result of CH1 and the green font is the result of CH2 Utility Press Menu Button to enter Menu Description Frequency Measure...

Page 19: ...to enter shortcut key settings menu and choose shortcut key function After setting press this button single time to respond to the corre sponding function Default settings Long press button to enter a...

Page 20: ...ack forms to the measured point The current value of the measured point will be displayed on the screen 3 Resistance measurement Press the power button to turn on then press the DMM button to enter to...

Page 21: ...required measuring gear correctly and then measure again This chapter introduces the function of wave form generator Operation description Press the power button to turn on and then press the AWG but...

Page 22: ...an be observed through an oscilloscope Output the sine waveform Output a sine waveform with 10KHz 2 5Vpp as follows 1 Press AWG button to enter the waveform generator function interface 2 Press F1 to...

Page 23: ...les Produkte JT DMSO2D72 DMSO2D72_Software zip 2 Install the driver Connect the oscilloscope to the computer via the USB cable Open the device manager of the computer and find the device If the device...

Page 24: ...signal output 5 Select signal type as arb and set the corresponding frequency and amplitude 6 Select Arb Channel as Arb1 Arb2 Arb3 Arb4 Each arb channel can save only one arbitrary waveform which was...

Page 25: ...automatically stop charging Storage and Replacement of the battery Storage Lithium batteries must be stored in clean dry and ventilated rooms Contact with corrosive substances should be avoided and th...

Page 26: ...ls to output Turn up the magnitude for DC signals with large magnitude In addition you may press the Auto button to perform an automatic detection of signals at first Contact Joy IT Technical Support...

Page 27: ...do not use any corrosive or chemical cleaning agents All specifications herein mentioned apply to the series oscilloscopes Before check ing an oscilloscope to see if it complies with these specificat...

Page 28: ...d accumulators that are not enclosed by waste equipment must be separated from it Return options As an end user you can return your old device which essentially performs the same function as the new d...

Reviews: