Joerns Dolphin Fluid Immersion Simulation System User & Service Manual Download Page 9

9

Joerns® Advanced Support Surfaces

Dolphin Fluid Immersion Simulation® Systems

© 2014 Joerns Healthcare • 6110109 RevG • 14-2640

In the 

Bed

 position, the system operates normally, 

but when switched to the 

Chair/Stretcher

 position, 

the timing cycles change to allow use on a specialty 

surface.

Using the specialty surface allows the system to be 

moved from the patient bed to a wheelchair cushion 

or stretcher pad, providing the normal functions of the 

system.

Alarm (Figure 4)

The warning or alarm subsystem consists of LED’s 

and a beeper which displays red and beeps when a 

fault condition occurs.

A fault condition is considered to be any of the 

following conditions: 

•  Pressure too hard for more than a 10 second period

•  Pressure too soft for more than a 10 minute period

•  Differential error between “

Comfort Adjust

 setting” 

and “

Auto Feedback

” for more than a 30 minute 

period

The beeper may be manually disabled for up to 30 

minutes by pressing the yellow 

Alarm

 button.

This feature avoids annoyance while a fault is being 

corrected, but will automatically re-assert itself after 30 

minutes time, or until the fault is corrected. The LED’s 

continue to function normally, regardless of the 

Alarm

 

on/off state.

Lock (Figure 4)

The 

Lock

 button and associated yellow LED permit 

the entire control panel to be locked from further 

adjustments. 

When locked, pressing the 

Lock

 button again restores 

normal operation and the yellow LED is extinguished.

Battery Indicators (Figure 4)

The Battery indicator will blink when the AC power has 

been interrupted and the control unit is running on the 

battery back-up power. 

The 

Battery Low

 indicator will blink when the battery 

back-up is at the end of its charge life. Plug control 

unit back into a power outlet as soon as possible to 

resume normal operation. Upon restoration of AC 

power, the battery back-up will begin the recharge 

process. 

Note:

 To ensure the battery recharges when 

connected to AC power, the 

Storage Switch

 must be in 

the 

Battery On

 position.

Figure 6

Figure 5

Immersion Profile Window (Figure 5)

The 

Immersion Profile

 indicates the system response 

to patient initial positioning and position change. When 

in optimal position, the green LED will illuminate. 

When the system is in transition, the yellow/red LEDs 

will illuminate. The Dolphin FIS System will recreate 

the optimal profile based on individual patient body 

characteristics. 

No

 

manual adjustment is needed

.

Comfort Controls (Figure 6)

Autofirm

The 

Autofirm

 mode is strictly used for patient transfers, 

repositioning and to quickly inflate the surface when it 

has not been in use. 

Caution:

 The System should never be left in 

Autofirm

 

mode while a patient is on the surface outside of 

transfers or repositioning. 

Autofirm

 is not a therapy 

mode.

To override the 

Autofirm

 mode, press the 

Autofirm

 

button again.

Summary of Contents for Dolphin Fluid Immersion Simulation System

Page 1: ...dvanced Surfaces Dolphin Fluid Immersion Simulation System Dolphin Pediatric Fluid Immersion Simulation System To avoid injury read user s manual before using Representative image of the Dolphin Fluid...

Page 2: ...y within the confines of the bed Do not let any extremities protrude over the side or between the bed rails when the mattress is being used Danger Explosion Hazard Do not use in the presence of flamma...

Page 3: ...for assessment and implementation of bed side rails in various health care settings Entrapment zones involve the relationship of bed components often directly assembled by the healthcare facility rath...

Page 4: ...pecialty Surface Wheelchair Cushion Troubleshooting 12 Therapy Surface Not Inflating Unable to Change Therapy Mode or Adjust Comfort Control Nursing Procedures 12 13 Recommended Linen Changing the The...

Page 5: ...on FIS System is an advanced therapy system designed to provide state of the art pressure redistribution by simulating the effects of a body immersed in a fluid medium The Dolphin System includes thre...

Page 6: ...bution technology designed to alleviate vertical shear forces The therapy mattresses and specialty surfaces conform to the specific shape of the patient minimizing soft tissue distortion reducing isch...

Page 7: ...ess that can fit any healthcare bed frame including bariatric frames up to 48 122 cm wide Conforms to specific shape of the patient minimizing soft tissue distortion reducing ischial tuberosity penetr...

Page 8: ...battery back up will provide power for continued therapy mattress system operation for approximately 12 hours Keyboard Functions Warning For important precautions see page two Caution The patient s he...

Page 9: ...d from further adjustments When locked pressing the Lock button again restores normal operation and the yellow LED is extinguished Battery Indicators Figure 4 The Battery indicator will blink when the...

Page 10: ...0 LED s and cover the full control range from Soft to Firm When operating within normal parameters the Auto Feedback LED scale will be amber Should the system be outside of normal parameters the LED s...

Page 11: ...e appropriate position based on the patient s clinical condition 2 Select the highest or most firm Comfort Adjust setting 3 Hand Check Place a hand with three 3 fingers if head of bed at 30 or higher...

Page 12: ...g on Battery Back up Ensure that the power is not on Standby If on Standby press the Power button Ensure that all air cells are connected Ensure that the Mode is in the appropriate position for the at...

Page 13: ...that the patient is positioned properly within the confines of the bed or other specialty item Do not let any extremities protrude over the side or between the bed rails when the mattress is being use...

Page 14: ...he therapy mattress or specialty surface deflated thoroughly wipe down entire mattress as air cells will lie flat Be sure to reach all areas underneath and in between air cells Allow to air dry If dus...

Page 15: ...retcher Pad Maximum weight capacity2 700 lbs 318 1 Kg Wheelchair Cushion Maximum weight capacity3 250 lbs 113 6 Kg Pediatric System Minimum weight capacity1 5 lbs 2 3 Kg Dimensions Control Unit 11 5 2...

Page 16: ...should be observed to verify normal operation in the configuration in which it will be used Guidance and Manufacturer s Declaration Electromagnetic Emissions The AutoVector Fluid Immersion Simulator...

Page 17: ...ospital environment Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital env...

Page 18: ...ation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference...

Page 19: ...ng to the maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz...

Page 20: ...e Improper maintenance handling and or cleaning Failure to use in the manner indicated in the Dolphin FIS System user manual Any modification repair or alteration done to the Dolphin FIS System that w...

Reviews: