Guidance and Manufacturer’s Declaration – Electromagnetic Immunity

The AutoVector

®

 Fluid Immersion Simulator is intended for use in the electromagnetic environment specified below. 

The customer or the user of the Auto Vector Fluid Immersion Simulator should assure that it is used in such an 

environment. 
Immunity test

IEC 60601 test level Compliance level Electromagnetic environment – guidance

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3 

3 Vrms

150 kHz to 80 MHz

3 V/m

80 MHz to 2,5 GHz

3 Vrms

3 V/m 

Portable and mobile RF communications equipment 

should be used no closer to any part of the Auto Vector 

Fluid Immersion Simulator, including cables, than the 

recommended separation distance calculated from the 

equation applicable to the frequency of the transmitter.

Recommended separation distance

d=1.2√P

 

d=1.2√P

  80 MHz to 800 MHz

d=2.3√P

  800 MHz to 2,5 GHz

Where 

P

 is the maximum output power rating of the 

transmitter in watts (W) according to the transmitter 

manufacturer and 

d

 is the recommended separation 

distance in metres (m). 

Field strengths from fixed RF transmitters, as 

determined by an electromagnetic site survey, 

a

should 

be less than the compliance level in each frequency 

range.

 b

 

Interference may occur in the vicinity of equipment 

marked with the following symbol: 

Note 1: 

At 80 MHz and 800 MHz, the higher frequency range applies.

Note 2: 

These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 

reflection from structures, objects and people.

a

  Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land 

mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically 

with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site 

survey should be considered. If the measured field strength in the location in which the Auto Vector Fluid Immersion 

Simulator is used exceeds the applicable RF compliance level above, the Auto Vector Fluid Immersion Simulator 

should be observed to verify normal operation. If abnormal performance is observed, additional measures may be 

necessary, such as re-orienting or relocating the Auto Vector Fluid Immersion Simulator.

b

   Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Summary of Contents for Dolphin Fluid Immersion Simulation System

Page 1: ...dvanced Surfaces Dolphin Fluid Immersion Simulation System Dolphin Pediatric Fluid Immersion Simulation System To avoid injury read user s manual before using Representative image of the Dolphin Fluid...

Page 2: ...y within the confines of the bed Do not let any extremities protrude over the side or between the bed rails when the mattress is being used Danger Explosion Hazard Do not use in the presence of flamma...

Page 3: ...for assessment and implementation of bed side rails in various health care settings Entrapment zones involve the relationship of bed components often directly assembled by the healthcare facility rath...

Page 4: ...pecialty Surface Wheelchair Cushion Troubleshooting 12 Therapy Surface Not Inflating Unable to Change Therapy Mode or Adjust Comfort Control Nursing Procedures 12 13 Recommended Linen Changing the The...

Page 5: ...on FIS System is an advanced therapy system designed to provide state of the art pressure redistribution by simulating the effects of a body immersed in a fluid medium The Dolphin System includes thre...

Page 6: ...bution technology designed to alleviate vertical shear forces The therapy mattresses and specialty surfaces conform to the specific shape of the patient minimizing soft tissue distortion reducing isch...

Page 7: ...ess that can fit any healthcare bed frame including bariatric frames up to 48 122 cm wide Conforms to specific shape of the patient minimizing soft tissue distortion reducing ischial tuberosity penetr...

Page 8: ...battery back up will provide power for continued therapy mattress system operation for approximately 12 hours Keyboard Functions Warning For important precautions see page two Caution The patient s he...

Page 9: ...d from further adjustments When locked pressing the Lock button again restores normal operation and the yellow LED is extinguished Battery Indicators Figure 4 The Battery indicator will blink when the...

Page 10: ...0 LED s and cover the full control range from Soft to Firm When operating within normal parameters the Auto Feedback LED scale will be amber Should the system be outside of normal parameters the LED s...

Page 11: ...e appropriate position based on the patient s clinical condition 2 Select the highest or most firm Comfort Adjust setting 3 Hand Check Place a hand with three 3 fingers if head of bed at 30 or higher...

Page 12: ...g on Battery Back up Ensure that the power is not on Standby If on Standby press the Power button Ensure that all air cells are connected Ensure that the Mode is in the appropriate position for the at...

Page 13: ...that the patient is positioned properly within the confines of the bed or other specialty item Do not let any extremities protrude over the side or between the bed rails when the mattress is being use...

Page 14: ...he therapy mattress or specialty surface deflated thoroughly wipe down entire mattress as air cells will lie flat Be sure to reach all areas underneath and in between air cells Allow to air dry If dus...

Page 15: ...retcher Pad Maximum weight capacity2 700 lbs 318 1 Kg Wheelchair Cushion Maximum weight capacity3 250 lbs 113 6 Kg Pediatric System Minimum weight capacity1 5 lbs 2 3 Kg Dimensions Control Unit 11 5 2...

Page 16: ...should be observed to verify normal operation in the configuration in which it will be used Guidance and Manufacturer s Declaration Electromagnetic Emissions The AutoVector Fluid Immersion Simulator...

Page 17: ...ospital environment Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital env...

Page 18: ...ation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference...

Page 19: ...ng to the maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz...

Page 20: ...e Improper maintenance handling and or cleaning Failure to use in the manner indicated in the Dolphin FIS System user manual Any modification repair or alteration done to the Dolphin FIS System that w...

Reviews: