The 

Autofirm 

button causes the therapy mattress or 

specialty surface to fill to maximum inflation. After 15 

minutes, the system will automatically reset to the 

previous inflation level.

While in the 

Autofirm

 mode, the 

Comfort Adjust

 

indicator

 

LED will remain on its normal setting to show 

where the inflation will return upon resumption of 

normal operation. Also, the 

Comfort Adjust

 indicator 

will blink amber at the firm position when in 

Autofirm

 

mode. There is no restriction against the user 

immediately returning to the 

Autofirm

 mode once 

leaving that mode. 

The 

Comfort Adjust

 indicator indicates where the 

manual pressure adjustment is set by the 

Comfort 

Adjust 

arrows

.

Comfort Adjust

Joerns recommends that caregivers allow 

the Dolphin FIS System to set and control the 

immersion profile.

 However, to accommodate 

individual patient preference, caregivers may use the 

Comfort Adjust

 arrows to manually adjust comfort 

settings. It is recommended that manual adjustments 

of more than one (1) LED step up or down from the 

system profile be avoided.

Note:

 The Dolphin FIS System automatically adjusts 

the neutral buoyant immersion profile based on 

individual patient characteristics.  The 

Comfort 

Adjust

 feature is designed to allow for individual 

patient comfort preferences.  Should the patient 

request adjustment due to bed articulation, such as 

head of bed elevation, this may be accomplished by 

increasing the 

Comfort Adjust

 indicator up or down 

incrementally one LED.  Care should be taken to 

minimize adjustments and allow the System to control 

the therapy surface’s optimal profile.

Auto Feedback

The 

Auto Feedback

 indicator scale is represented by 

10 LED’s and cover the full control range from 

Soft

 to 

Firm

When operating within normal parameters, the 

Auto 

Feedback

 LED scale will be amber. Should the system 

be outside of normal parameters, the LED scale will 

move from amber to red, indicating a potential need to 

manually adjust with 

Comfort Adjust

 arrows.

It is normal for the 

Auto Feedback

 LED to move to 

red when the patient is transitioning on the therapy 

mattress. Allow the Dolphin FIS System to optimize. If 

the LED lights remain consistently red after the system 

has had the chance to optimize, manual adjustment 

with the 

Comfort Adjust

 arrows is needed.

Grounding Instructions

Warning:

 Use a properly grounded, AC outlet for this 

product. Failure to use a grounded outlet could result 

in personal injury or damage to equipment or structure 

wiring, including risk of fire. A qualified electrician 

should be contacted to correct the wiring and ensure a 

properly grounded outlet. 

Before installing this product, have the electrical 

system checked to make sure the electrical circuits 

and the electrical service are properly grounded. There 

is always a chance of a loose connection or poor 

installation of a ground wire that causes the loss of 

proper ground at the outlet. Inadequate grounding at 

electrical outlets can occur even if there is a ground 

wire. Wires can become loose over time at the 

connection to the outlet. 

Note:

 To install new wires on 

a circuit requires a qualified electrician. 

How to Determine if Your Outlet has the 
Proper Grounding 

Most hardware stores sell circuit testers that can 

be used to test an outlet for proper grounding. The 

tester plugs into an outlet and by observing the 

indicator lights you can determine if the outlet is 

properly grounded. For a higher level of assurance, an 

electrician should be requested to thoroughly test the 

electrical system with more reliable equipment. 

If repair or replacement of the cord or plug is 

necessary, please contact Joerns Healthcare for 

assistance.

Summary of Contents for Dolphin Fluid Immersion Simulation System

Page 1: ...dvanced Surfaces Dolphin Fluid Immersion Simulation System Dolphin Pediatric Fluid Immersion Simulation System To avoid injury read user s manual before using Representative image of the Dolphin Fluid...

Page 2: ...y within the confines of the bed Do not let any extremities protrude over the side or between the bed rails when the mattress is being used Danger Explosion Hazard Do not use in the presence of flamma...

Page 3: ...for assessment and implementation of bed side rails in various health care settings Entrapment zones involve the relationship of bed components often directly assembled by the healthcare facility rath...

Page 4: ...pecialty Surface Wheelchair Cushion Troubleshooting 12 Therapy Surface Not Inflating Unable to Change Therapy Mode or Adjust Comfort Control Nursing Procedures 12 13 Recommended Linen Changing the The...

Page 5: ...on FIS System is an advanced therapy system designed to provide state of the art pressure redistribution by simulating the effects of a body immersed in a fluid medium The Dolphin System includes thre...

Page 6: ...bution technology designed to alleviate vertical shear forces The therapy mattresses and specialty surfaces conform to the specific shape of the patient minimizing soft tissue distortion reducing isch...

Page 7: ...ess that can fit any healthcare bed frame including bariatric frames up to 48 122 cm wide Conforms to specific shape of the patient minimizing soft tissue distortion reducing ischial tuberosity penetr...

Page 8: ...battery back up will provide power for continued therapy mattress system operation for approximately 12 hours Keyboard Functions Warning For important precautions see page two Caution The patient s he...

Page 9: ...d from further adjustments When locked pressing the Lock button again restores normal operation and the yellow LED is extinguished Battery Indicators Figure 4 The Battery indicator will blink when the...

Page 10: ...0 LED s and cover the full control range from Soft to Firm When operating within normal parameters the Auto Feedback LED scale will be amber Should the system be outside of normal parameters the LED s...

Page 11: ...e appropriate position based on the patient s clinical condition 2 Select the highest or most firm Comfort Adjust setting 3 Hand Check Place a hand with three 3 fingers if head of bed at 30 or higher...

Page 12: ...g on Battery Back up Ensure that the power is not on Standby If on Standby press the Power button Ensure that all air cells are connected Ensure that the Mode is in the appropriate position for the at...

Page 13: ...that the patient is positioned properly within the confines of the bed or other specialty item Do not let any extremities protrude over the side or between the bed rails when the mattress is being use...

Page 14: ...he therapy mattress or specialty surface deflated thoroughly wipe down entire mattress as air cells will lie flat Be sure to reach all areas underneath and in between air cells Allow to air dry If dus...

Page 15: ...retcher Pad Maximum weight capacity2 700 lbs 318 1 Kg Wheelchair Cushion Maximum weight capacity3 250 lbs 113 6 Kg Pediatric System Minimum weight capacity1 5 lbs 2 3 Kg Dimensions Control Unit 11 5 2...

Page 16: ...should be observed to verify normal operation in the configuration in which it will be used Guidance and Manufacturer s Declaration Electromagnetic Emissions The AutoVector Fluid Immersion Simulator...

Page 17: ...ospital environment Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital env...

Page 18: ...ation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference...

Page 19: ...ng to the maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz...

Page 20: ...e Improper maintenance handling and or cleaning Failure to use in the manner indicated in the Dolphin FIS System user manual Any modification repair or alteration done to the Dolphin FIS System that w...

Reviews: