Joerns® Camtec Series Bed Frames
RC 1000 Bed
© 2016 Joerns Healthcare • 6110284 RevB • 16-2016
Periodic Inspection and Maintenance
(See Customer Maintenance Check List)
The amount of maintenance required by the bed will be dictated by its use. As a minimum the unit should be
periodically inspected every 6 months.
1. After a week or two of usage, check all the bolts to be sure that none of them have worked loose and that all pins
are in their normal location and securely fastened. Check all welds.
2. At least once every six months the actuator drive screw should be lubricated using a heavy duty grease such as
Drydene #5871 or equivalent.
3. Put light machine oil on all pivot points every 6 months.
4.
Check all electrical connections for tightness.
5. Check all electrical wiring for fraying, kinking damage and/or deterioration.
IF ANY DISCREPANCIES ARE NOTED DURING INSPECTION, THEY MUST BE CORRECTED BEFORE
CONTINUED USE OF THE BED.
Joerns Healthcare Technical Support can provide additional assistance to repair or troubleshoot any problems.
Major service and repair must only be performed by authorized Joerns Healthcare Technical Support, which you may
contact at 800.966.6662.
Troubleshooting
Problem
Possible Cause
Solution
No functions operate
No power to unit
Check that the power cord is plugged
into receptacle and light on control
module is illuminated.
Check for cord breaks
Power light does not light
Electrical cords not properly plugged
into control module
Check all electrical cords
and connections.
Power light is on but no functions
operate
Bad pendant
Replace pendant
Bad control module
Replace control module
Three motors operate “OK” One
motor will not operate
Bad pendant
Isolate the Problem
Interchange motor connections on
control module. Swap connections
on the module so that problem motor
is plugged into a motor connection
on the module that is working. If the
problem motor now operates and
the motor that was OK now does not
operate, it indicates either the control
module or the pendant is faulty.
Replace as required.
Bad control module
Bad motor
Bad wiring
If the above does not resolve the problems, call Joerns Healthcare Technical Support at 800.966.6662. Have the
serial number of the bed and make the phone call from as close to the bed as practical.