![Joerns Camtec RC1000 User & Service Manual Download Page 10](http://html1.mh-extra.com/html/joerns/camtec-rc1000/camtec-rc1000_user-and-service-manual_2021304010.webp)
10
Joerns® Camtec Series Bed Frames
RC 1000 Bed
© 2016 Joerns Healthcare • 6110284 RevB • 16-2016
Customer Maintenance Check List
S/N Bed____________________________________ S/N Controller__________________________________
Bed Model__________________________________ Date_____________________
Initial Date
1
Verify what options are on the bed:
Foot End Side Rails
Scale
Head End Side Rails
Trapeze
Pads (If Applicable)
I.V. Pole
Other______________________________________________________________
2
Verify and record the S/N
3
Check Head Section operation with pendant
4
Check Knee Gatch operation with pendant
5
Check the Bed Elevation operation with pendant
6
Check the Trendelenburg operation with pendant (If Applicable)
7
Check the Seat operation with the pendant (If Applicable)
8
Check casters function: Steer________ Total________
9
Inspect the hardware to ensure that components are fit securely
10 Check to ensure that the Electrical components are secure
11 Check Power Cord for the following:
A: Check all strain reliefs and cable clamps are secure
B: Check for worn areas, cuts or frays
C: Check plug for bent or loose prongs
12 Check that Pendant cord has no damage and that it is secured properly to the control box
13 Check to ensure that all decals are affixed at proper location
Head, Foot and Motor decals
Power and Control Cord on Head Board
Power and Control on the Control Box
14 Check the presence and fit of the Mattress Stop
15 Check the presence, fit and operation of the Emergency Crank
16 Check the operations of all accessory options noted in #1 above
17 Check to ensure that the side rails function properly and lock in place
18 Measure and record the Leakage Current value should be less than 250 microamperes
(UL 60601-1) Use Calibrated Model LT544D-20 Digital Safety Analyzer or equivalent
Measured Value____________
19 Measure and record Line Grounding Resistance. Value should read equal to or less than 0.1
OHMS as measured between the ground wire in the line cord and the bed frame.
Measured Value____________
20 Visually inspect the bed for paint or assembly defects; touch up as needed
21 Check the Top Expansion (If Applicable) of the Mattress during operation at:
Head Section
Right ________
Left ________
Center Section
Right ________
Left ________
Upper Knee Gatch
Right ________
Left ________
Lower Knee Gatch
Right ________
Left ________