Loss of Power
If it becomes necessary to disconnect the bed power while the scale is in use:
1. Note the displayed patient weight before disconnecting the power.
2. Unplug the bed and move bed and patient to the new location.
3. When the patient is settled restore the power to the bed.
4.
Press the “
FREEZE/RESUME
” button to WAKE-UP the readout. If the weight reading is correct, the system
is ready.
5. If the displayed reading is not correct:
a.
Press the “
FREEZE/RESUME
” button. The display will read “
HOLD (weight)
”
b.
Use the “
WEIGHT
” and “
WT/CHG
” keys to ramp the displayed value to the correct weight.
c. When done, press the “
FREEZE/RESUME
” button once again to save the correction and exit to the normal
weighing mode. If you make a mistake, press “
ON
” to about your changes.
Important:
If you do not know the correct patient weight, you must remove the patient from the bed and re-zero
the system.
1. Press the “
FREEZE/RESUME
” button. The display will read: “
FREEZE = (patient weight)
”.
2. Make all necessary adjustments to bed and/or patient.
3. Press the “
FREEZE/RESUME
” button once again to resume normal weight monitoring.
Moving Bed and Patient
If it becomes necessary to move the bed and/or patient while the scale system is active, use the following procedure:
Toggles between POUNDS and KILOGRAMS display modes.
While in the
FREEZE
mode, the “
WEIGHT
” and “
WT/CHG
” keys will allow you to manually adjust the displayed weight
reading in the event of a know error. Simply press the “
WEIGHT
” and “
WT/CHG
” keys to make the proper corrections.
Note:
If
FREEZE
mode is left unattended for more than three minutes, display will read “
LEFT IN FREEZE - REFER
TO MANUAL
”. You must press the “
FREEZE/RESUME
” button to exit this mode.
This feature will allow you to make adjustments to the bed without effecting the displayed
patient weight.
1. Press the “
FREEZE/RESUME
” button. The display will read: “
FREEZE = (patient weight)
”.
2. Make all necessary adjustments to the bed.