![Joerns Camtec RC1000 User & Service Manual Download Page 8](http://html1.mh-extra.com/html/joerns/camtec-rc1000/camtec-rc1000_user-and-service-manual_2021304008.webp)
8
Joerns® Camtec Series Bed Frames
RC 1000 Bed
© 2016 Joerns Healthcare • 6110284 RevB • 16-2016
Five Function Hand Pendant Operation
The
HEAD
up and down buttons cause the head of the bed to raise and lower.
The
FOOT
up and down buttons cause the knees and lower portion of the legs (Knee
Gatch) to be raised and lowered.
The
BED
up and down buttons cause the height of the bed to raise and lower.
The
SEAT
up button causes the bed to go into a modified cardiac chair position. The
head section and head end of the bed will raise, the knee section will raise and the
foot end of the bed will lower.
The
SEAT
down button will cause the bed to return to a supine position (CPR
Position)
The
TREND
down button will cause the foot end of the bed to raise and the head end
to lower with a limiting angle of 12 degrees.
The
TREND
up button will cause the foot end of the bed to go down and the head
end of the bed to raise until reaching a limit of 15 degrees.
The bed contains four electric drive modules. A hand crank is also supplied, at the head end of the bed; which can be
used to adjust the bed in the case of a power or actuator failure. Disconnect the power before using the crank. The
pendant should not be used during hand cranking.
Adjusting the Bed Width
Caution:
Extend all four sections of each side
prior
to adjusting bed position or
side rails.
1. Raise the bed to the highest position
2. Pull down on the blue lanyard to disengage the locking pin. A slight inward or
outward pressure on the expansion pan may be required for ease in disengaging
the locking pin.
3. Move adjustable mattress pan section about 1" in desired direction.
4.
Release the blue lanyard.
5. Move adjustable mattress pan section in desired direction until locking pin snaps
into locking hole.
6. Repeat this procedure until all adjustable mattress pan sections are at the
desired locations.
Cleaning and Maintenance
Cleaning
Warning:
High pressure cleaning or power wash are not to be used on any frame part.
Warning:
Lock casters when cleaning, inspecting or performing any maintenance on the bed. Lower all mattress
support platform sections unless necessary for service access.
Before cleaning any portion of the bed, disconnect the power cord from the wall outlet
1. Using standard cleaning procedures, wipe down the frame of the bed. Be sure to wear appropriate protection for
the type of materials being used.
2. Use cleaning detergents sparingly around the pendant controller, electronic controller and the motors.