Important Precautions 

Caution: 

Rotating parts. Keep fingers and arms clear 

of bed when mechanism is moving.

Caution: 

Pinch points. Keep objects away from these 

areas while bed is in motion.

Caution:

 Electrical hazard may occur if bed is plugged 

into inadequate power supply. Make sure to use a 115 

volt, 3 prong, grounded outlet with a capacity of at 

least 15 amps.

Caution:

 Keep feet clear of casters. Make sure 

casters are locked before transferring patients.

Caution:

 Make sure that the cords from the control 

module and pendant are properly positioned before 

operating the bed. Do not allow either cord to get 

caught in the moving parts of the bed.

Caution:

 When using a turning mattress, make sure 

there is a bolster between the mattress and side rail.

Warning:

 Keep all objects out from under bed. Certain 

parts under the bed raise and lower when the bed is 

 

in motion.

Warning:

 Do not place arms or head into frame 

while operating. Stand clear of the bed frame before 

operating.

Warning:

 Do not operate this device if the power cord 

or any of the cords coming from the control module are 

cut, frayed, damaged or loosely connected.

Warning:

 Do not service the bed without first 

disconnecting the power cord.

Warning:

 The bed must be unplugged from the wall 

socket prior to using the emergency crank. Failure to 

do so may result in damage to the bed and injury to 

the personnel using the crank.

Warning:

 ESD - electrostatic discharge.

Warning:

 The bed should be left in the lowest position 

when unattended in order to reduce the risk of injury 

due to falls while getting into or out of bed, or while 

lying on the bed.

Warning:

 When raising or lowering bed, casters at 

the foot end should be in the steer position. Head end 

should be free to move.

Warning: Possible Fire Hazard.

 Use nasal mask or 

½ bed tent oxygen administering equipment. Oxygen 

tent should not extend below mattress support 

platform. Pendant should not be placed in an oxygen 

enriched environment such as an oxygen tent. Use of 

electrical circuits in an oxygen-enriched environment 

could result in a fire hazard.

Warning:

 

Possible Injury Or Death.

 An optimal 

bed system assessment should be conducted on 

each resident by a qualified clinician or medical 

provider to ensure maximum safety of the resident. 

The assessment should be conducted within the 

context of, and in compliance with, the state and 

federal guidelines related to the use of restraints 

and bed system entrapment guidance, including 

the 

Clinical Guidance for the Assessment and 

Implementation of Side Rails

 published by the Hospital 

Bed Safety Workgroup of the U.S. Food and Drug 

Administration. Further information can be obtained 

at the following web address: http://www.fda.gov/

MedicalDevices/ProductsandMedicalProcedures/

GeneralHospitalDevicesandSupplies/HospitalBeds/

default.htm.

Warning: Possible Injury Or Death.

 Use a properly 

sized mattress in order to minimize the gap between 

the side of mattress and Side Rails/Assist Devices. 

This gap must be small enough to prevent resident/ 

patient from getting his/her head or neck caught in 

this location. Make sure that raising or lowering bed, 

or contouring the mattress support platform, does 

not create any hazardous gaps. Excessive gaps may 

result in injury or death.

Warning: Possible Injury Or Death.

 Residents/ 

patients may become entangled in pendant cord. 

Patients with reduced mental acuity should not be 

allowed access to pendant. Unsupervised use of 

pendant could result in injury or death.

Caution:

 Before transporting bed, unplug bed from 

power source, secure Power Cord on cord storage 

straps on Head Panel and hang Pendant from Head 

Panel. Failure to do so could result in Power Cord, 

pendant cord, or bed damage, thereby creating a 

potential hazard.

Caution: 

Extend all four sections on each side prior to 

adjusting bed position or side rails. 

Warning: 

Use Width Extenders only after the resident/ 

patient is safely positioned in the bed. Failure to do so 

could result in injury.

Summary of Contents for Camtec RC1000

Page 1: ...User Service Manual Joerns Camtec Series Bed Frames RC 1000 Bed Expandable Split frame Bariatric Bed Model RC1000 To avoid injury read user s manual before using The Caregiver s Choice TM...

Page 2: ...nt should not extend below mattress support platform Pendant should not be placed in an oxygen enriched environment such as an oxygen tent Use of electrical circuits in an oxygen enriched environment...

Page 3: ...group HBSW The workgroup s goal is to improve the safety of bed frames for patients and residents in all health care settings who are most vulnerable to the risk of entrapment The efforts of the FDA a...

Page 4: ...low it to pass through a 40 door opening The frame is constructed of structural steel The frames lowest position is approximately 16 from the floor to the mattress support platform The frame in the hi...

Page 5: ...30 to 95 Non Condensing Atmospheric Pressure 700 hPa to 1060 hPa Storage and Shipping Conditions Ambient Temperature 30 C to 50 C Relative Humidity 10 to 95 Atmospheric Pressure 500 hPa to 1060 hPa Be...

Page 6: ...two halves as indicated 8 Transfer the head and foot end to their new location Keep the head and foot section matched together per serial number ID labels Transfer the head board and foot board to the...

Page 7: ...e when raising or lowering the bed damage to the bed may result 2 Plug the power cord into a grounded 115 volt socket having a capacity of at least 15 amps Make sure light on control module is illumin...

Page 8: ...t the power before using the crank The pendant should not be used during hand cranking Adjusting the Bed Width Caution Extend all four sections of each side prior to adjusting bed position or side rai...

Page 9: ...ajor service and repair must only be performed by authorized Joerns Healthcare Technical Support which you may contact at 800 966 6662 Troubleshooting Problem Possible Cause Solution No functions oper...

Page 10: ...rays C Check plug for bent or loose prongs 12 Check that Pendant cord has no damage and that it is secured properly to the control box 13 Check to ensure that all decals are affixed at proper location...

Page 11: ...Lubricate the actuator drive screws using a heavy duty grease such as Drydene 5871 or equivalent This should be done approx every 6 months 24 If bed is a Scale Bed check that the scale functions are o...

Page 12: ...12 Joerns Camtec Series Bed Frames RC 1000 Bed 2016 Joerns Healthcare 6110284 RevB 16 2016 Appendix A Wiring Diagram...

Page 13: ...13 Joerns Camtec Series Bed Frames RC 1000 Bed 2016 Joerns Healthcare 6110284 RevB 16 2016 Appendix A Scale System Assembly...

Page 14: ...14 Joerns Camtec Series Bed Frames RC 1000 Bed 2016 Joerns Healthcare 6110284 RevB 16 2016 Appendix A Main Frame Assembly...

Page 15: ...15 Joerns Camtec Series Bed Frames RC 1000 Bed 2016 Joerns Healthcare 6110284 RevB 16 2016 Appendix A Mattress Support Assembly...

Page 16: ...16 Joerns Camtec Series Bed Frames RC 1000 Bed 2016 Joerns Healthcare 6110284 RevB 16 2016 Appendix A Caster Base Assembly...

Page 17: ...17 Joerns Camtec Series Bed Frames RC 1000 Bed 2016 Joerns Healthcare 6110284 RevB 16 2016 Appendix A Label Location...

Page 18: ...18 Joerns Camtec Series Bed Frames RC 1000 Bed 2016 Joerns Healthcare 6110284 RevB 16 2016 Appendix A Labels Used on Bed...

Page 19: ...16 x 1 LG 17 902518 0 Nut Hex 16 Elastic Lock 18 300591 0 Assembly Foot Board Support 19 300592 0 Assembly Head Board Support 20 106167 0 Emergency Crank Clamp Bracket 21 106166 0 Emergency Crank Bra...

Page 20: ...2 58 903809 0 Cable Hi Lo Foot End Motor Split to Control Box 59 903812 0 Hi Lo Head End Actuator 60 903803 0 Hi Lo Foot End Actuator 61 104592 0 Spacer x 9 16 62 901007 0 Screw 10 32 x Round Head Ph...

Page 21: ...ease Plunger 98 904342 0 Spring Compression 99 904343 0 Caplug COF 100 900016 0 Nut Hex 20 Elastic Lock 101 900258 S Washer Lock Split SS 102 900991 0 Washer Nylon x x 1 16 103 901071 0 Screw 20 x LG...

Page 22: ...e ID RC750 RC1000 Left Foot 131 904447 0 Label Slide ID RC750 RC1000 Right Foot 132 904448 0 Label Slide ID RC750 RC1000 Mid Section 133 904449 0 Label Slide ID RC750 RC1000 Upper Knee Gatch 134 90390...

Page 23: ...000 0 LB The patient may now be placed on the bed Important When the 0 0 button is pressed everything on the weighing surface will be zeroed out on the display Make sure that all linens pillows etc a...

Page 24: ...weight you must remove the patient from the bed and re zero the system 1 Press the FREEZE RESUME button The display will read FREEZE patient weight 2 Make all necessary adjustments to bed and or patie...

Page 25: ...and Disinfecting The readout and transducer for the 1228 Scale System are made of aluminum and steel with a painted surface You need only exercise caution when cleaning the display window as this is...

Page 26: ...zed in writing by Joerns Healthcare will void this warranty Parts Joerns beds contain various parts that wear from normal use These parts are not covered under the stated warranty but are covered for...