To Remove Control Box

1.  Plugs and outlets are numbered and color-coded 

so they can be easily reinstalled later (See 

Actuator And Control Box Identification).

2.  Unplug from wall outlet.

3.  Unplug all actuator cords and pendant/staff  

control plug.

4. 

Remove the control box retaining clip (Figure 33).

5. 

Slide the Control Box off the actuator assembly.

6. 

To replace the Control Box, reverse 

 

these instructions.

To Remove Actuator
Recommended Tools: Screwdriver, Hammer,  
Masking Tape, Zip Ties

1. 

Identify the actuator to be replaced.

2.  Tip the bed on its side to replace high/low 

actuator(s).

3.  Unplug the actuator cord.

4. 

The actuator is held in place by two pins. Use a 

screwdriver to remove the lock ring on each pin.

5. 

Slide the pins out of holes. If necessary, use the 

hammer to tap out the pins.

6. 

You may now replace the actuator(s).

7.  To reassemble the bed, reverse the previous 

steps, making note of the following two important 

items:

•  Assemble the pin and lock rings as  

originally installed.

• 

Zip-ties should be replaced, with cords returned 

to their original position and routing direction to 

the control box.

Figure 33

To Remove AC Power Cable
Recommended Tools: Small Screwdriver, Masking 
Tape, Zip Ties, 5/16" Nut Driver

1. 

Raise the bed so that there is about 27 inches 

of clearance between the floor and the mattress 

support platform of bed.

2.  Unplug the power cable from the outlet.

3. 

Cut the zip-ties that hold the power cable to the 

frame. Tape the remaining cables in their original 

position. It is very important to keep the actuator 

cables routed along the frame in their original 

factory-installed positions. If this step is not 

followed, damage to cables or electric shock may 

result.

4.  Detach the ground wire from the bed frame  

if equipped.

Warning: Possible Shock Hazard.

 Injury may result 

from improper power cable routing. Always follow 

proper factory-installed routing configuration. Be sure 

to check condition of the power cable on a regular 

basis. Do not use damaged or frayed power cables.

5. 

Using a small screwdriver, alternately depress 

the small red latch tabs above and below AC 

receptacle on the control box. Simultaneously, 

pull gently outward on the power cable where it 

connects to the control box, until the plug comes 

out.

6.  Plug the replacement cable into the control box 

until it latches. Attach the ground wire to the bed 

frame.

7. 

Re-route the power cable along the frame in the 

same position as the original routing.

8.  Replace zip-ties along with cables in their original 

position and routing to the control box. If this step 

is not followed, damage to the cables or electric 

shock may result.

Summary of Contents for EasyCare

Page 1: ...User Service Manual Joerns Bed Frames EasyCare Bed To avoid injury read user s manual before using...

Page 2: ...rom any outlet Warning Possible Shock Hazard Injury may result from improper routing of the power cord Always follow the proper factory installed routing configuration Failure to do so could result in...

Page 3: ...if mattress used with Mattress Side Stop does not fill the entire width between stops or which compresses to less than 1 5 inches under user s weight Warning Possible Injury or Death Floor locks incr...

Page 4: ...Therefore compliance is the responsibility of the facility As the leading manufacturer of long term care beds and a frontrunner in addressing this critical issue Joerns Healthcare can offer you the ex...

Page 5: ...tection LA27 Actuators Auto Reset Current Sensor Environmental Conditions Operating Conditions Ambient Temperature 10 C to 40 C Relative Humidity 30 to 75 Non Condensing Atmospheric Pressure 700 hPa t...

Page 6: ...ain support of the bed panel Insert the safety pins on each side and latch them Figure 3 6 Press the mounting bracket against the bed frame and tighten the four panel mounting screws right and left si...

Page 7: ...h 12 printed on it It is longer than the standard bumper 1 Remove the shorter bumper by tapping the tangs on either side of the leg with a rubber mallet 2 Place the 12 baseboard bumper 3 Tap into plac...

Page 8: ...3 Ensure that the cable plug is fully inserted into the cable jack 4 To secure tighten the union nut on both cables Figure 14 Operation Pendant Operation Your EasyCare bed comes standard with one of t...

Page 9: ...tion button Standard Staff Control Figure 21 2 1 4 6 5 3 7 Advanced Staff Control Figure 22 8 2 12 7 1 4 6 5 3 9 10 11 Head Operation 1 Push to raise Back 2 Push to lower Back Knee Operation 3 Push to...

Page 10: ...Install correct baseboard bumper Bed Transport Roll in Low Lock System 1 To keep the bed from rolling when in the lowest position press down on the locking lever to lock casters Figure 23 2 The stand...

Page 11: ...re 28 and 29 5 Release the tether to engage the detent Figure 27 6 Repeat the above steps on the remaining MSP extensions Warning Possible Injury Or Death Use a mattress that is properly sized to fit...

Page 12: ...in intact and are securely fastened to their intended connection point Actuator and Control Box Information Warning Possible Shock Hazard Unplug the power cord from the wall outlet before performing a...

Page 13: ...iver Masking Tape Zip Ties 5 16 Nut Driver 1 Raise the bed so that there is about 27 inches of clearance between the floor and the mattress support platform of bed 2 Unplug the power cable from the ou...

Page 14: ...k of use check all threaded fasteners for looseness and make sure all pins are in their normal location and fastened securely Check monthly for loose bolts nuts pins and other retaining hardware Tight...

Page 15: ...the floor and or cause the bed to become unstable resulting in injury or death Drainage Bag Holder Warning Possible BioHazard Prior to use consult the Drain Bag manufacturer s installation and use in...

Page 16: ...igure 36 1 The Drainage Bag Holder is designed as an attachment point for items such as urine collection bags Make sure items attached to holder are safely secured Prior to use please read important w...

Page 17: ...be fully charged Note We recommend that users charge up the Battery immediately upon receipt 4 The Battery has a useful life of about 3 5 years depending upon usage It is recommended to discharge the...

Page 18: ...are covered for 90 days after date of delivery Joerns Healthcare s obligation under this warranty is limited to supplying replacement parts servicing or replacing at its option any product which is fo...

Reviews: