background image

Joerns

Healthcare

Joerns Healthcare 1_Art Direction_Joerns Beds_700 Series_UltraCare 8page
5/2007_Page 1 of 8

User-Service Manual

Joerns

®

 Support Surface

DermaFloat

®

 APM

transforming wound management

To avoid injury, read user’s manual before using.

Summary of Contents for DermaFloat APM

Page 1: ...Joerns Healthcare User Service Manual Joerns Support Surface DermaFloat APM transforming wound management To avoid injury read user s manual before using...

Page 2: ...presence of flammable anesthetics Do not use in the presence of smoking materials or open flame Air flowing through the air mattress will support combustion Danger To reduce the risk of shock adhere...

Page 3: ...esponsibility of the facility As the leading manufacturer of long term care beds and a frontrunner in addressing this critical issue Joerns Healthcare can offer you the expertise assistance and produc...

Page 4: ...when equipment is placed on bed frames that leave gaps of even a few inches between the mattress and the head panel foot panel and bed or side rails The equipment is NOT to be used when such gaps are...

Page 5: ...ts should be considered on an individual basis and discussed with the attending physician Features The DermaFloat APM is comprised of two components Therapy control unit Therapy mattress system Therap...

Page 6: ...though there is no ground wire There is always a chance of a loose connection or poor installation of a ground wire that causes the loss of proper ground at the outlet Inadequate grounding at electric...

Page 7: ...de Operation Warning For important precautions see page two Caution The resident s head should be positioned in the center of the top section of the mattress When using the mattress system always ensu...

Page 8: ...ll vary based on resident weight and profile To resume therapy reattach the hose to the control unit Figure 3 CairLock The mattress system comes equipped with our specially designed CairLock therapy c...

Page 9: ...or the highest comfort control setting to achieve quick inflation Locate the Ready Valve by unsnapping the flap on the hose set Turn the Ready Valve to inflate or deflate as desired Return the control...

Page 10: ...s of resident migration or inadvertent deflation of resident surface Skin Care Monitor skin conditions regularly particularly in areas where incontinence and drainage occur or collect and consider adj...

Page 11: ...hands Therapy Pad The therapy pad can be wiped down with a disinfectant solution or a mild detergent with a damp cloth If heavily soiled the therapy pad can be laundered in a washer and dryer with wa...

Page 12: ...ght must not exceed bed frame manufacturers specified load capacity Environmental Conditions Operating Conditions Ambient temperature 10 C to 40 C Relative humidity Non condensing Storage And Shipping...

Page 13: ...13 Joerns Support Surfaces DermaFloat APM Notes...

Page 14: ...essive usage Improper maintenance handling and or cleaning Failure to use in the manner indicated in the DermaFloat APM user manual Any modification repair or alteration done to the DermaFloat APM tha...

Reviews: