•  Resident care equipment that is used in isolation 

areas should be disinfected in accordance with all 

internal policies and procedures regarding such 

equipment.

Disinfecting

When there is visible soilage and between residents, 

we recommend that the unit and mattress be 

disinfected with a tuberculocidal disinfectant. 

Disinfectant should be registered with the 

Environmental Protection Agency (EPA).

•  Use rubber gloves and eye protection. 

•  Prepare detergent/disinfectant (registered by EPA 

as hospital disinfectant) solution according to 

instructions on label for correct use-dilution. 

•  With support surface deflated, thoroughly wipe down 

entire mattress, as air cells will lie flat. Be sure to 

reach all areas underneath and in-between air cells. 

Allow to air dry. 

•  If dust or other soiling has accumulated along 

air hoses, remove using swabs moistened with 

detergent/disinfectant as necessary. Allow all 

components to air dry. Wrap mattress in plastic and 

return to storage area. 

•  Thoroughly wipe down outside of control unit and 

allow to air dry. Cover with plastic and return to 

storage area.

•  Remove gloves and dispose; wash hands.

 

Therapy Pad

The therapy pad can be wiped down with a disinfectant 

solution or a mild detergent with a damp cloth. If 

heavily soiled, the therapy pad can be laundered in a 

washer and dryer with warm water (no more than 120º 

Fahrenheit). A non-bleach detergent should be used 

sparingly. Wipe dry or allow to air dry.

Steam Cleaning

Do not use any steam cleaning device on the unit. 

Excessive moisture can damage mechanisms in  

this unit.

Maintenance

Warning: 

Only facility-authorized personnel trained 

by Joerns should perform preventative maintenance. 

Preventative maintenance performed by unauthorized 

personnel could result in personal injury or equipment 

damage. Any maintenance done without Joerns’s 

authorization will invalidate any warranties on this 

product.

Storage and Care

When the product is not in use, properly store  

the power cord. Failure to do so could result in 

personal injury.

Note: 

Clean the DermaFloat

®

 APM as described in the 

previous section prior to storage.

Control Unit

The power cord may be wrapped around the unit for 

convenience. Wrap the unit in a plastic bag for dust 

resistance, then store the unit in an area appropriate 

for an electronic medical device. 

Support Surface

Gently roll up the support surface, expelling any 

residual air, for temporary storage. The mattress 

should be wrapped in plastic and/or a clean bag   

for storage. 

Summary of Contents for DermaFloat APM

Page 1: ...Joerns Healthcare User Service Manual Joerns Support Surface DermaFloat APM transforming wound management To avoid injury read user s manual before using...

Page 2: ...presence of flammable anesthetics Do not use in the presence of smoking materials or open flame Air flowing through the air mattress will support combustion Danger To reduce the risk of shock adhere...

Page 3: ...esponsibility of the facility As the leading manufacturer of long term care beds and a frontrunner in addressing this critical issue Joerns Healthcare can offer you the expertise assistance and produc...

Page 4: ...when equipment is placed on bed frames that leave gaps of even a few inches between the mattress and the head panel foot panel and bed or side rails The equipment is NOT to be used when such gaps are...

Page 5: ...ts should be considered on an individual basis and discussed with the attending physician Features The DermaFloat APM is comprised of two components Therapy control unit Therapy mattress system Therap...

Page 6: ...though there is no ground wire There is always a chance of a loose connection or poor installation of a ground wire that causes the loss of proper ground at the outlet Inadequate grounding at electric...

Page 7: ...de Operation Warning For important precautions see page two Caution The resident s head should be positioned in the center of the top section of the mattress When using the mattress system always ensu...

Page 8: ...ll vary based on resident weight and profile To resume therapy reattach the hose to the control unit Figure 3 CairLock The mattress system comes equipped with our specially designed CairLock therapy c...

Page 9: ...or the highest comfort control setting to achieve quick inflation Locate the Ready Valve by unsnapping the flap on the hose set Turn the Ready Valve to inflate or deflate as desired Return the control...

Page 10: ...s of resident migration or inadvertent deflation of resident surface Skin Care Monitor skin conditions regularly particularly in areas where incontinence and drainage occur or collect and consider adj...

Page 11: ...hands Therapy Pad The therapy pad can be wiped down with a disinfectant solution or a mild detergent with a damp cloth If heavily soiled the therapy pad can be laundered in a washer and dryer with wa...

Page 12: ...ght must not exceed bed frame manufacturers specified load capacity Environmental Conditions Operating Conditions Ambient temperature 10 C to 40 C Relative humidity Non condensing Storage And Shipping...

Page 13: ...13 Joerns Support Surfaces DermaFloat APM Notes...

Page 14: ...essive usage Improper maintenance handling and or cleaning Failure to use in the manner indicated in the DermaFloat APM user manual Any modification repair or alteration done to the DermaFloat APM tha...

Reviews: