49/55
Dx is the value of D measured[F].
De is the accuracy of D.
6.1.23 Rs Accuracy
When Dx (measured D value) ≤ 0.1
Rs accuracy is given by the following:
Rse = Xx*De
[Ω]
Xx = 2πfLx=1divided by 2πfCx
Here, Xx is the value of the measured X.
[S].
Cx is the value of the measured C [F].
Lx is the value of L measured[H].
De is the accuracy of D.
F is test frequency
6.1.24 Accuracy factor
Figure A, basic measurement accuracy A
In figure A, a smaller value is selected on the boundary line.
In figure A, the basic accuracy A value selection method is as follows: