32
Symptom
Assumed reason
Countermeasure
Refer
Unable to display
saved image.
File corrupted.
(E002)
Unable to restore corrupted file. Please
delete the corrupted file.
p.23
Incorrect calendar
information
shown with saved
file
Calendar
information inside
the instrument is
incorrect.
Clock may off sets if you do not use the
instrument for a while. Reconfigure the
clock.
Unable to charge a
battery pack
Battery pack is
deteriorated
The battery pack is deteriorated if it does
not get charged fully. The battery pack is
expendable accessory. Please purchase
another designated battery pack.
-
Charging out of
charger’s
operational
temperature range
The battery charger’s operational
temperature is between 0
℃
and 40
℃
.
Please charge in the above mentioned
temperature range.
-
Continuous
operation time is
short even with
fully charged
battery pack
The temperature
inside the instrument
is high
Turn the power OFF or bring it over to
an environment in low temperature to
bring the temperature of battery pack
itself. Once the temperature of the
battery pack itself becomes low, its
operation returns normal.
Continuous use time length becomes
shorter as the temperature inside the
instrument becomes higher. This is due
to battery pack’s protection circuit
becomes active, and not malfunction of
the battery pack.
-
Battery pack is
deteriorated
The battery pack is deteriorated if its
operation time length is short under
normal temperature range. The battery
pack is expendable accessory. Please
purchase another designated battery
pack.
-