substances the liberation of hazardous gases may require the use of 
a fume cupboard or other means of extraction. 

 

Be  aware  of  biological  contamination.  Ensure  sterilization  of  the 
capillary  tube,  waste,  nebuliser  and  mixing  chamber  is  performed 
after contamination.  

 

Ensure  equipment  is  used  on  a  clean,  dry,  non-combustible,  solid 
work surface with at least 300mm suitable clearance all around and 
at least 1m clearance above the chimney. 

 

Ensure that the outer chimney is properly installed before operating 
the equipment. 

 

Attention: 

Ensure  that  the  gas  and  air  pipes  have  been  connected 

correctly and that there are 

no gas leaks

 present before igniting. 

 

Ensure  that  the  unit  is  properly  connected  to  the  gas  cylinder 
making  sure  there  are  no  gas  leaks.  Ensure  that  the  correct  gas 
regulator  is  used  and  connected  properly.  The  responsible  body 
must  ensure  that  the  type  of  connector  used  on  the  outlet  of  the 
gas-pressure  regulator  conforms  to  the  applicable  national 
requirements. 

 

Take  care  when  fitting  and  removing  the  capillary  tube  to  the 
nebulizer needle. 

 

It  is  not  recommended  to  leave  the  flame  photometer  unattended 
during operation. 

 

Do not 

operate or handle any part of the product with wet hands.  

 

Do not 

look down the flame chimney when igniting or operating the 

equipment. 

 

Do not 

handle the flame chimney whilst in operation. 

 

Do not 

lean or stretch over the equipment whilst in operation. 

 

Do  not 

cover  the  chimney  whilst  in  use. 

Do  not

  block  or  obstruct 

ventilation slots / airways.

 

 

Do  not 

spill  substances  onto  the  unit.  If  spillage  does  occur, 

disconnect  unit  from  mains  supply.  If  using  hazardous  substances 
please adhere to standard laboratory operating procedures.

 

 

Do not

 position the product so that it is difficult to disconnect from 

the mains supply.  

 

Do  not

  move  or  carry  the  unit  when  in  use  or  connected  to  the 

mains electricity supply. 

 

Important

:  Allow  sufficient  time  for  the  chimney  to  cool  before 

handling. 

 

Do not

 use acetylene with the flame photometer. 

 

Warning: 

Over adjustment of the fuel valve will cause excess flame.  

DOMINIQUE DUTSCHER SAS

Summary of Contents for PFP7

Page 1: ...500 795 Rev D 07 14 Flame Photometer Models PFP7 and PFP7 C Operating Manual D O M I N I Q U E D U T S C H E R S A S...

Page 2: ...ny intended operation The fault condition should immediately be reported to the appropriate servicing authority NOTE your basic safety protection afforded by the equipment and may invalidate the warra...

Page 3: ...lator conforms to the applicable national requirements Take care when fitting and removing the capillary tube to the nebulizer needle It is not recommended to leave the flame photometer unattended dur...

Page 4: ...he manufacturer responsible bodies should check with the manufacturer that the proposed method will not damage the equipment In order to completely disconnect power to the flame photometer the mains p...

Page 5: ...de laboratoire 5 En cas de suspicion d alt ration de la s curit de l appareil de quelque mani re que ce soit celui ci doit tre mis hors service et son utilisation doit tre rendue impossible L anomalie...

Page 6: ...un d gagement d au moins 1m au dessus de la chemin e Assurez vous que la chemin e ext rieure soit correctement install e avant d utiliser l quipement Attention assurez vous que les tuyaux de gaz et d...

Page 7: ...vertissement un r glage inappropri du robinet de combustible permettant un d bit excessif entrainera une flamme trop importante vitez d installer le photom tre de flamme dans une zone expos e aux cour...

Page 8: ...zzati Implementare le procedure di laboratorio comunemente accettate per la manipolazione sicura dei prodotti chimici 5 Se si sospetta che la sicurezza dell unit sia stata compromessa disattivare e re...

Page 9: ...sopra la ciminiera Assicurarsi che la ciminiera esterna sia stata adeguatamente installata prima di utilizzare l apparecchiatura Attenzione Prima dell accensione assicurarsi che i tubi del gas e dell...

Page 10: ...sovraregolazione della valvola del carburante genera una fiamma eccessiva Non installare il fotometro a fiamma in zone soggette a correnti ma assicurarsi che vi sia una ventilazione adeguata per evit...

Page 11: ...ceptados para la manipulaci n segura de los productos qu micos 5 Si se sospecha que la seguridad de la unidad ha sido alterada de alguna forma la unidad debe inutilizarse y deben tomarse medidas para...

Page 12: ...y al menos 1 m de espacio libre por encima de la chimenea Aseg rese de que la chimenea externa est instalada apropiadamente antes de poner en funcionamiento el equipo Atenci n Aseg rese de que las tub...

Page 13: ...e excesivo de la v lvula de combustible provocar una llama excesiva Evite instalar el fot metro de llama en zonas expuestas a corrientes de aire pero aseg rese de que haya una ventilaci n adecuada par...

Page 14: ...Es sind allgemein anerkannte Laborverfahren f r die sichere Handhabung von Chemikalien einzusetzen 5 Falls vermutet wird dass die Sicherheit des Ger ts auf irgendeine Weise beeintr chtigt wurde ist d...

Page 15: ...n jede Richtung befindet und mindestens 1 m Platz oberhalb des Schornsteins vorhanden ist Stellen Sie sicher dass der u ere Schornstein vor dem Betrieb des Ger ts ordnungsgem installiert wurde Achtung...

Page 16: ...lammenphotometer Achtung Eine berm ige Justierung des Brennstoffventils f hrt zu berm iger Flammenbildung Stellen Sie das Flammenphotometer nicht in zugigen Bereichen auf aber sorgen Sie daf r dass ei...

Page 17: ...lysis Preparation 28 3 1 Calibration standards 28 3 2 Sample preparation 30 3 3 Dilution 31 SECTION 4 32 Operation 32 4 1 Front and rear panel controls 32 4 2 Operation Flame Ignition and Optimisation...

Page 18: ...e results 44 7 3 Unable to set standard reading 44 7 4 Non linear results 45 7 5 Flame will not light 45 7 6 No electrical power 45 7 7 Unable to set blank 46 7 8 No reading on display 46 SECTION 8 46...

Page 19: ...ed 0 10 0 mmol L K Limits of Detection Na 0 2ppm Na 0 2ppm K 0 2ppm K 0 2ppm Li 0 25ppm Li 0 25ppm Ca 15ppm Ba 30ppm Reproducibility 1 Coefficient of variation C V for 20 consecutive samples using 10p...

Page 20: ...oncentration the intensity of the emission is directly proportional to the number of atoms returning to the ground state This is in turn proportional to the absolute quantity of the species volatized...

Page 21: ...avelengths are easily isolated from those of most other elements due to wide spectral separation The analysis of Na K Li Ba and Ca are typically determined at low temperatures i e 1500 2000 C therefor...

Page 22: ...ing standards The performance of the instrument depends upon the accuracy and purity of the calibration standards 4 If standard solutions are required to be stored for any length of time or at an elev...

Page 23: ...2 L P G regulated to 10 15 inches water gauge 0 36 0 54psi or 0 025 0 038kg cm2 Suitable regulators are available from Jenway refer to Section 2 5 2 1 3 Air A supply of dry clean and pulse free air at...

Page 24: ...onal as ordered 013 046 Nebuliser inlet tube 500mm 500 193 Gas tube low pressure 2 metres 500 191 Silicon rubber tube drain 023 003 Air tube 2 metres 500 192 Nozzle 1 gas 026 012 Nozzle 2 air 026 013...

Page 25: ...he restrictor screw 5 fitted into the end of the fuel inlet 6 on the mixing chamber The restrictor screw should remain in place if butane propane or L P G is to be used as fuel Keep the restrictor scr...

Page 26: ...Fuel to mixing chamber inlet 7 Mixing chamber to drain outlet 8 End cap 9 Drain trap clip 10 Drain trap to waste containter tubing 11 Mixing chamber to drain trap tubing 12 Compressed air to nebuliser...

Page 27: ...tor 2 Plug the power cable into the POWER IN socket 17 figure 2 3 2 If necessary connect a suitable plug to the 3 pin AC supply lead The colours of the wires conform to the internationally recognised...

Page 28: ...or variations in specification that could degrade the performance of the instrument Accessories Part Code Model 8515 air compressor 220V 50Hz 535 001 Model 8516 air compressor 110V 60Hz 535 002 Butane...

Page 29: ...2 Always ensure that the standards encompass the expected range of the sample concentrations 3 Standards should be prepared so as to ensure that the region in which measurements are made coincide wit...

Page 30: ...olution for use with the flame photometer this stock solution should be diluted 1 in 50 Calculation Atomic weight K 39 1 Molecular weight KCl 74 56 Therefore 0 477g KCl contains 0 477 x 39 1 0 25g K 7...

Page 31: ...oration of the solvent that could lead to elevated concentrations 6 When in doubt about the equipment or application the operator should contact Jenway at jenwayhelp bibby scientific com for advice 3...

Page 32: ...us levels of dilution will often be necessary Good quality deionised water should normally be used for carrying out these dilutions and it is recommended that the same batch of water should be used fo...

Page 33: ...led by the front panel power switch 2 Ignition A spring loaded switch which when depressed will cause an electrical discharge between the ignition electrode and the burner unit thereby causing fuel ig...

Page 34: ...itions 3 and 4 will normally be blank 4 1 2 Rear panel controls Figure 4 1 2 1 Rear panel 1 VOLTAGE SELECTION A two position switch marked 230 and 115 These positions allow operation from the voltage...

Page 35: ...or 0 025 0 38kg cm2 See section 2 1 2 for information on different fuel types 7 AIR 5 16 connector for air tubing IMPORTANT Air pressure on the inlet to the instrument must not exceed 30psi 2kg cm2 Se...

Page 36: ...irate a standard solution of slightly higher concentration than expected in samples to be tested Adjust the fine and coarse control until a positive reading obtained This is a nominal value to be used...

Page 37: ...o that they lie on the linear part of the curve The user can manually construct a calibration curve using standard solutions containing known concentrations of the elements to be determined and if nec...

Page 38: ...ck the blank setting and adjust Aspirate the remaining standard solutions if used to construct the calibration curve note the When the blank and standards are set unknown samples can be aspirated and...

Page 39: ...least 1 in 100 and set the display reading accordingly using the fine coarse sensitivity 4 Adjust the fuel valve in a clockwise direction until a peak reading is obtained Re set display reading accord...

Page 40: ...me is alight 4 The top of the instrument chimney unit becomes very hot when running and can cause severe burns if touched 5 The exhaust gases from the flame are very hot and the area approximately 1 m...

Page 41: ...6 Finger operated plungers for easy draining 5 1 2 Specifications Model Large Small Part code 500 177 500 176 Port size Maximum operating pressure 100psi 100psi Working temperature range 5 to 80 C wit...

Page 42: ...ply to the flame photometer is switched on 2 Connect the air tubing supplied with the flame photometer to the compressor outlet nozzle and secure with a hose clip 3 Connect the other end of the air tu...

Page 43: ...r one minute 3 The consumption rate should be between 2 and 6ml min If correct no further action is required 4 If the consumption rate is too low this is likely to be caused by a blockage in the fine...

Page 44: ...llow 30 minutes for the burner system to cool down 3 Disconnect the mixing chamber from the chimney unit by unscrewing the knurled locking ring and dropping the chamber clear of the chimney 4 Remove t...

Page 45: ...n 4 2 Operation Fuel cylinder nearly empty Replace cylinder with a fully charged one Instrument in strong draught Relocate the instrument Instrument in very bright sunlight Relocate the instrument Ins...

Page 46: ...e temperature not correctly set Ensure the fuel valve is correctly set for the element being determined refer to Section 4 2 Instrument not draining adequately the drain trap is at the correct level 7...

Page 47: ...1 Minor Spares Kit PFP7 MINKIT Description Part code Quantity Nebuliser I 500019 1 Cleaning Wire 3x150mm I 500194 1 Cleaning Solution 025016 2 Ignition Probe Assembly PFP7 IPA 1 Mk2 Small Knob Blank...

Page 48: ...500 795 Rev D 07 14 Page 47 D O M I N I Q U E D U T S C H E R S A S...

Page 49: ...500 795 Rev D 07 14 Page 48 D O M I N I Q U E D U T S C H E R S A S...

Page 50: ...33 1 64 45 13 00 Fax 44 0 1785 810405 e mail info bibby scientific com e mail bsf bibby scientific fr e mail bibby bibby scientificasia com www bibby scientific com www bibby scientific com www bibby...

Reviews: