background image

 

 

Safety

 

Please read this information carefully prior to installing or using this equipment. 

1.  The 63 series spectrophotometers are designed for operation by 

trained personnel

 who are 

aware of the principles and applications involved. For further help and advice please visit 

www.jenway.com

 

 

2.  The spectrophotometer is a sensitive electronic and optical instrument designed for use in a 

laboratory environment. Careful adherence to the installation instructions must be observed. 
If in doubt contact a 

relevant and competent authority

 for advice before proceeding. 

 

3.  In addition to observing the instructions detailed in the Operating Manual and Service Manual 

for this instrument all installation, operating and service personnel must be aware of, and 
employ, 

a safe system of work

.  

 

4.  Voltage levels hazardous to life are present in this instrument, for personal safety only 

trained engineers

 aware of the risk and avoidance of electric shock should remove 

protective covers from the instrument. 

 

5.  This instrument is designed for minimal maintenance, which must be carried out carefully 

following the 

procedures detailed in this manual

. All safety instructions in these 

procedures, as well as those defined locally for the 

area or environment

 where the work is 

being carried out must be observed. 

 

6.  Other than for those items defined in the maintenance procedures herein there are 

no user 

serviceable

 items in this instrument. Removal of covers and attempted adjustment or service 

by unqualified personnel will invalidate any warranty and incur additional charges for repair. 

 

7.  Reference should always be made to the 

Health and Safety Data

 for any chemicals or 

reagents used.  All available information, advice and warnings on the handling, storage, use 
and disposal of such must be carefully observed. When not available this data must be 
requested from the supplier before proceeding in any way.  

 

8.  It is important that 

good laboratory practice

 is observed when handling samples, chemicals, 

reagents and ancillary equipment in order to carry out measurement and analysis with this 
instrument. Suitable 

safety and personal protective equipment

 must be used at all times. 

 

9.  If it is suspected that safety protection has been impaired in any way, the spectrophotometer 

must be made 

inoperative and secured

 against any intended operation. The fault condition 

must be reported to the 

appropriate servicing authority

. In all such reports the 

model 

number and serial number

 of the spectrophotometer must be quoted. 

 

 

Sécurité 

 
Ces informations sont importantes ; veuillez les lire avec attention avant d'installer ou 
d'utiliser l'instrument. 

 

1.  Les spectrophotomètres de la série 63 sont conçus pour être utilisés par du

 personnel 

formé

 qui connaît les principes et les applications concernés. Pour obtenir de l'aide et des 

conseils supplémentaires, merci de consulter le site Internet 

www.jenway.fr

  

 

 

2.  Le spectrophotomètre est un instrument électronique et optique sensible conçu pour être 

utilisé en laboratoire. Les instructions d'installation doivent être scrupuleusement respectées. 

Summary of Contents for 6305

Page 1: ...Operating Manual 635 015 REV B 06 14 Model 6305 Spectrophotometer...

Page 2: ...2...

Page 3: ...val of covers and attempted adjustment or service by unqualified personnel will invalidate any warranty and incur additional charges for repair 7 Reference should always be made to the Health and Safe...

Page 4: ...n de ces produits chimiques doivent tre respect s soigneusement Lorsqu elles ne sont pas disponibles ces donn es doivent tre demand es dans tous les cas au fournisseur avant de poursuivre 8 Il est imp...

Page 5: ...eine geeignete Sicherheitsausr stung und pers nliche Schutzausr stung zu verwenden 9 Falls vermutet wird dass der Sicherheitsschutz auf irgendeine Weise beeintr chtigt wurde ist das Spektrophotometer...

Page 6: ...are o utilizzare lo strumento 1 Gli spettrofotometri della serie 63 sono progettati per essere utilizzati da personale qualificato che conosce i principi e le applicazioni ad essi collegati Per ulteri...

Page 7: ...tiche di laboratorio quando si manipolano campioni sostanze chimiche reagenti e apparecchiature accessorie per eseguire misurazioni e analisi con questo strumento Attrezzature idonee per la protezione...

Page 8: ...RICS MEASUREMENTS 18 4 1 TRANSMISSION MEASUREMENT MODE 18 4 2 ABSORBANCE MEASUREMENT MODE 18 SECTION 5 CONCENTRATION MEASUREMENTS 20 5 1 USING A KNOWN STANDARD 20 5 2 USING A KNOWN FACTOR 20 SECTION 6...

Page 9: ...ontamination Combined with a mechanically rigid structure this model provides a system with fast warm up low drift and high reliability 1 2 INSTRUMENT SPECIFICATION 6305 Wavelength Range 198 to 1000nm...

Page 10: ...he mains fuse is housed within the POWER IN socket When replacing the fuse the unit must be disconnected from the mains supply The fuse rating is 2A F fast blow type In the event of the fuse failing a...

Page 11: ...g 2 3 1 Display 1 Primary display area Transmission Absorbance Concentration 2 Primary display adjust annunciator 3 Secondary display area Wavelength Factor 4 Primary display units 5 Secondary display...

Page 12: ...u option 4 CAL Key Initiates a calibration routine 5 Print Key Provides a printout of the current reading with an incremental sample number When pressed for the first time after a calibration the prin...

Page 13: ...ess panel and all ventilation slots must NOT be covered or obstructed at any time 2 Power switch On off switch for the unit 3 Power in socket IEC type connection socket for mains cable 4 Output socket...

Page 14: ...is travelling through The Beer Lambert law states that A is the absorbance is the molar absorption coefficient l mol 1 cm 1 c is the concentration mol l 1 l is the path length cm This law shows that...

Page 15: ...enon lamp is focused onto the grating with 1200 lines per millimeter which separates the light into discreet wavelengths The diffracted spectrum of light then passes through a further slit and lens ar...

Page 16: ...to 260nm Below this level quartz cuvettes must be used NOTE the plastic cuvettes supplied with the unit are not suitable for use below a wavelength of 320nm b Plastic disposable cuvettes should only b...

Page 17: ...In the preparation of samples and standards high grade borosilicate glass and AR grade chemicals and reagents must be used Good quality deionised water or other suitable solvents must be used for diss...

Page 18: ...screen Subsequent samples can be measured in the same way If the wavelength is adjusted between sample measurements then the instrument must be calibrated again before more samples can be measured 4 2...

Page 19: ...19 measured...

Page 20: ...ously Use the arrow keys to change the concentration to the required value for the standard Press the CAL or RETURN key to calibrate the instrument to the standard concentration The instrument will ca...

Page 21: ...also be sent to the 6305 via the serial interface using Windows HyperTerminal for example ASCII D or d Same as pressing the PRINT key ASCII T CR Outputs transmission and wavelength separated by an AS...

Page 22: ...readily incorporated into most spreadsheet software packages 6 1 RS232 Output The bi directional RS232 interface is available on the rear panel 25 way D type connector The connections are as follows T...

Page 23: ...PC GND 7 5 GND Note the interface cable kit 542 009 can be used to implement the above interconnections 6 2 Recorder Output This is available via the 4mm rear panel sockets The level is proportional...

Page 24: ...or use with computer without a serial port 543 001 40 Column printer supplied with interface cable paper roll and ribbon and universal power supply 060 287 Paper roll for printer quantity 1 060 288 Pr...

Page 25: ...or commercial use and will give rise to errors if used 8 2 LAMP REPLACEMENT 8 2 1 Xenon Flash Lamp Module Replacement Before replacing the lamp disconnect the unit from the mains supply Access to the...

Page 26: ...26...

Page 27: ...e Department for a reference number which you should include with your faulty equipment Please also ensure you include a clear description of the fault and a completed copy of our Decontamination Cert...

Page 28: ...is outside the instrument range Recalibrate the instrument using a standard of lower concentration Err 5 Lamp Failure The instrument is unable to detect a peak signal for wavelength calibration Possi...

Page 29: ...29 SECTION 10 Declaration of Conformity...

Page 30: ...748 Fax 33 1 64 45 13 00 Fax 44 0 1785 810405 e mail info bibby scientific com e mail bsf bibby scientific fr e mail bibby bibby scientificasia com www bibby scientific com www bibby scientific com ww...

Reviews: