22
calibrated. The shutter must be open for normal measurements.
ASCII
Gnnn<CR>
Commands the instrument to go to wavelength
nnm
. For example:
G540<CR> will set the wavelength to 540nm.
ASCII
Fxxxx.x<CR>
Sets the concentration factor to xxxx.x. For example: F1000<CR> will set the
factor to 1000.
Note
<CR>
is an ASCII carriage return character. The last three commands provide an output which
can be readily incorporated into most spreadsheet software packages.
6.1
RS232 Output
The bi-directional RS232 interface is available on the rear panel 25 way D type connector. The
connections are as follows:
TDX 2
– INPUT TO 6305
RXD 3
– OUTPUT FROM 6305
RTS 4
– LINKED TO CTS
CTS 5
– LINKED TO RTS
DSR 6
– OUTPUT FROM 6305
DCD 8
– OUTPUT FROM 6305
DTR 20
– INPUT TO 6305 (must be active)
GND 7
Suggested interconnections are detailed below:
6305
IBM PC XT (25 way “D”)
TDX 2
2 TXD (From PC)
RXD 3
3 RXD (To PC)
RTS 4
4 RTS (From PC)
CTS 5
5 CTS (To PC)
DSR 6
6 DSR (To PC)
DCD 8
8 DCD (To PC)
DTR 20
20 DTR (From PC)
GND 7
7 GND