Error processing
50
LDM52 OEM, Release 003, Revision 000
8
Error processing
In case of errors or when a measured value cannot be determined or put out, a warning or an
error message will be displayed.
Error
Meaning
Action
e1001
Error operation system/ firmware
Restart system
e1002
Error operation system/ firmware
e1003
Error operation system/ firmware
e1101
Error in communication with PC
check connection to external sys-
tem/ contact service for repair
e1102
Error in communication with PC
e1103
Laser module error
contact service/ send LDM52 OEM
to supplier
e1104
Laser module error
e1105
Laser module error
e1106
Hardware error (internal data transfer)
e1107
Hardware error (internal data transfer)
e1108
Hardware error (controller)
e1109
Hardware error (controller)
e1201
Measurement impossible / no target
adjust device/ check target
e1202
Error measuring module (calibration)
contact service/ send LDM52 OEM
to supplier
e1203
Target with unsuitable reflectivity
check target and distance
e1204
Measurement interrupted (measuring module)
Restart system
e1205
Measurement still running (measuring module)
Restart system
e1206
Target too bright / too much back light
check/ limit ambient light
e1207
Target outside of the measurement window (MW)
no action or change MW setup
e1208
Incorrect measurement parameterization
check setup
e1209
Hardware error (measuring module)
contact service/ send LDM52 OEM
to supplier
e1210
Current of laser is over the limit (laser stops work)
Restart of measurement (e.g. DT)
e1211
Stop of measurement (internal error)
Restart of measurement (e.g. DT)
Warnings
Meaning
Action
w1901
Restart being executed
no action
w1902
Input voltage outside of the specification (too low)
check power supply (10 … 30 VDC)
w1903
Input voltage outside of the specification (too high)
w1904
Temperature outside of the specification (too low)
check ambient temperature
w1905
Heating active, min. temperature not reached
no action/ wait until LDM52 OEM is
ready
w1906
Temperature outside of the specification (too high)
check ambient temperature
w1907
Trigger input and trigger output active at the same
time
Activate TRI or TRO, not both for the
same system
w1910
Measurement not completed within predefined pe-
riod of time
use variable measuring time (MF0)/
check target
w1911
Measuring frequency too high
change MF
w1912
Distance jump
Make sure a continuous movement
of target