Parameter setup and measuring operation
LDM52 OEM, Release 003, Revision 000
41
Value Description
Specification
x
Lower limit
x ≠ y
y
Upper limit
y ≠ x
Input of parameter QA in 0.1 mm.
Input:
QA10000 150000
Output: Parametrization of the analog switching output QA [QA]: 10000, 150000
The value of the output current (in mA) is calculated as follows:
x < y
mA
x
y
x
Dist
mA
mA
QA
16
4
]
[
Dist = measuring distance
Example 2:
A measuring range from 1 m up to 15 m shall be specified.
15 m should be the lowest current.
x = 15 m = 4 mA / y = 1 m = 20 mA
Value Description
Specification
x
Lower limit
x ≠ y
y
Upper limit
y ≠ x
Input of parameter QA in 0.1 mm
QA150000 10000
The value of the output current (in mA) is calculated as follows:
x > y
mA
y
x
y
Dist
mA
mA
QA
16
20
]
[
Dist = measuring distance
Entries of identical limits will be ignored and not accepted.