23
The Turbidity or Suspended Solids LINEAR analog output is based
on the following equation:
mA
(Turbidity or Suspended Solids)
= 4mA+(16mA)*(D
(Turbidity or Suspended Solids )
-
Turbidity or Suspended Solids (4) ) /Turbidity or Suspended Solids
(20) - Turbidity or Suspended Solids (4))
Where:
mA
(Turbidity or Suspended Solids)
= analog output
D
(Turbidity or Suspended Solids)
= current Turbidity or Suspended Solids display
Turbidity or Suspended Solids(4)
=Turbidity or Suspended Solids user setting for 4 mA
Turbidity or Suspended Solids(20)
= Turbidity or Suspended Solids user setting for 20 mA.