18
Note: The hysteresis value can be set within the range
corresponding to the measured electrode set. The user enters
the current setting screen after setting up the Hysteresis value.
e.
CURRENT OUT SETTING
page
1. 4mA OUT - User can
press [
] or [
] and
[
MODE/-
] key to enter the
desired values. Press the [
]
ENTER
key when finished.
(See chapter VI. 4-20mA
OUTPUT.)
2. 20mA OUT - Repeat the
same setting procedure of 4mA OUT for 20mA OUT setup. When
the 20mA OUT setup is complete, the user enters to the “WASH
SETTING”page. (See chapter VI. 4-20mA OUTPUT.)
f.
WASH SETTING
page
I
f the cleaning device is set to auto run, any one of the setting
values
reaches
to
zero
terminates this function. To
initiate the cleaning function,
the user can input the time in
seconds by using the [
],
[
] and [
MODE/-
] key. (The
default time is 0000 second
and range from 0000 to 9999
seconds). To terminate the
cleaning function, the user
can input the time in hours by
using the [
], [
] and
[
MODE/-
] key. (The default time is 000 hour and range from 000.0 to