Page 36
Shaker/Incubator Control Choices
In a case of the unit stops, the previous rpm, Temp,Timer
funcion can be used with slecting shaking and temp at the
same time
Press the START/SET button for
seven
times.”Snch”will
be
appeared on PV display and “on
the SV.
Use Up button and Down button
to set “ YES” or “ NO”
Press
Enter button to finish
the mode.
☞ When pressing START/STOP Button with “YES”on the SV
display, the Shaker/ Incubator operates at the same time.
Set “ No” on the SV display, The shaker doesn’ t operate.
☞ Even if setting “YES” on the SV display, the Shaker
function can be On/OFF with Shaker I/O Switch.
☞ Even if setting “No” on the SV display, the Shaker
function can be On/OFF with Shaker I/O Switch.
Shaker I/O Switch
This Switch is for controlling the Shaker On/Off without the
set value on the SV display. When pressing the switch once,
The shaker is on/off repeatedly once pressing the switch.
Able to check the set
value
modification
by
reoperation.