Page 47
(2) Electricity
Phenomenon
Check point
Settlement
Even if Power Switch
ON, OFF promptly..
Leak.
Stops a driving immediately and contacts
a reseller
A power plug being
Disconnected from an outlet.
Leaves Power Switch OFF, and please put a
power plug in an outlet.
Even if Power Switch
ON, an indication
does not work in
Display Panel.
Fuse Problem
Exchange the fuse positioned backside.
If temperature does
not rise
Confirms whether Temp Led
is turned on
If Temp Led is out, press Start/Stop
button.
If a machine stops
operation by Over Temp.
1 O/T LED is turned on
with red,
2. confirms whether a
Beep sound rings successively
1 If attributes red Knob of Over Temp.
Limit to a (-) driver clockwise and
presses one time of Start/Stop button
after currently adjusting than
temperature (PV) higher than 15%, red of
O/T Led disappears, and a Beep sound
disappears.
2. While Run LED and Heater LED are on if
they press one time of Start/Stop button
more, a machine operates again.
If the unit stops
operation
by Door Limit Switch
1. Door Led is turned on
with red
2.
Confirms
whether
a
Beep sound rings
1 Check whether Door LED is flickering
due to opened door
2. Confirms whether light reaches RUN LED
after having pressed one time of
START/STOP button.
Stops total control
without an indication
except the above
statement or button
operation.
1.Being affected by
strong Noise in a power
line
2. confirms whether the
high frequency welder
which strong high
frequency Noise occurs in
around appliances, mass
SCR Controller became
installation at hand
receive correct
acceptance.
Moves equipment to be the cause of, or a
change holds an Incubator installation
situation.