20
User ManUal
solvent refIll
1.
Turn OFF
the system and disconnect the power cord from the AC power source.
2.
Press the
cap lock
down and rotate
(counter-
clockwise)
the
refill cap
to expose the refill
ports.
Note:
You may hear a 'hiss' of air as the tank
depressurizes.
3.
Attach the
solvent refill cap assembly
to the
solvent bottle
.
4.
Push-in to attach the 2
quick connect valves
to the two refill ports, and
invert - hold
the
bottle until the system's cleaning solvent
reservoir is full.
5.
Press the
valve release
down to disconnect
the bottle and rotate the
refill cap
to original
position and turn the system ON.
6.
Prime
the system
(see page 16).
Cap Lock
2
3
solvent refill Cap
Assembly
Cleaning solvent Bottle
(see page 25 for solvent types)
4
5
Valve
release
Note:
Do not refill the solvent reservoir if the
LOW SOLVENT
light is not illuminated. Overfilling the tank can
prevent the system from pressurizing properly.
Depending on the type of cleaning tip used, you should expect
the following number of cleaning cycles per fUll solvent
reservoir.
2.5 mm tip
9,500 cycles
1.25 mm tip
12,500 cycles
ribbon/MtP tip
9,500 cycles
mAINTENANCE SChEDuLE
You must select one solvent type and use the
same solvent for the lifetime of the system.