CleanBlast – PortaBle
sYsteM statUs InDICators
This dual indicator will alert the user when there is AC
power connected to the system (red), and when the
nozzle safety mechanism is disengaged, allowing the
cleaning cycle to activate (green).
Note:
The
run button
on the handset
will not
operate when this LED is red
. To disengage the safety
mechanism, apply light pressure against the bulkhead
or patch cord and gently slide the handset forward.
When the POWER / READY LED turns GREEN, press the
RUN BUTTON on the handset to initiate cleaning.
POWER / READY
This indicator will alert the user that the level of
cleaning solvent
is too low for the system to operate
and will put the system into
standby mode.
The
solvent reservoir must be refilled before operation
can resume
(see page 20 for solvent refill procedures)
.
The reservoir will hold 225 ml. (8 fl. oz.) of solvent,
enough for at least 8,000 cleaning cycles.
Note:
Do not refill the solvent reservoir if the
LOW
SOLVENT
light is not illuminated. Overfilling the tank
can prevent the system from pressurizing properly.
LOW SOLvENT
solvent refill Cap
Assembly
Ordering Part Number
• FCLP-RCA-1
Quick Connect
Valves
Cleaning solvent Bottle
Ordering Part Numbers
• FCLP-SOL1
(1 bottle)
• FCLP-SOL1-6
(6 bottles)
This indicator will alert the user that the system
air filters
will need to be replaced. When the
reset
button
is pressed, the system cycle counter will turn
to zero, and the system will count to
100,000 cycles
before alerting the next filter change
(see page 21 for
air filter change procedures).
SERvICE
Air Filter
Ordering Part Number
• FCLP-FA-F1
(2 required per system)